Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Cureus ; 15(10): e47929, 2023 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-38046488

RESUMO

OBJECTIVE: The objective is to comparatively assess the impact strength and flexural strength of polyether ether ketone (PEEK) vs. computer-aided design/computer-aided manufacturing (CAD/CAM) polymethyl methacrylate denture base material. METHODS: A total of 90 samples were fabricated with traditional heat cure PMMA, PEEK, and CAD/CAM PMMA and divided into three groups of 30 samples each. The impact strength of all the samples was measured using an Izod impact tester with a pendulum in the air at 23±2°C. A three-point bending test was used in a Universal Testing Machine to assess the flexural strength of all the samples. The impact strength and flexural strength mean values were computed using a one-way ANOVA test. RESULT: Impact strength and flexural strength of PEEK (IS=10.22±1.25 kJ/m2 and FS=120±8.0 MPa) is almost identical to CAD/CAM PMMA sample (IS=9.595±3.313 kJ/m2 and FS=118.11±5.00 MPa) whereas for conventional heat cure PMMA (IS=4.00±.011 kJ/m2 and FS=75.4±4.50 MPa) the values are least among the three. CONCLUSION:  PEEK or CAD/CAM PMMA share almost identical and superior mechanical properties, and both can be used as better alternatives for complete denture fabrication rather than using conventional heat cure PMMA.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA