1.
Health Phys ; 108(4): 419-28, 2015 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-25706135

RESUMO

The authors have created a radiation transport code using the GEANT4 Monte Carlo toolkit to simulate pediatric patients undergoing CT examinations. The focus of this paper is to validate their simulation with real-world physical dosimetry measurements using two independent techniques. Exposure measurements were made with a standard 100-mm CT pencil ionization chamber, and absorbed doses were also measured using optically stimulated luminescent (OSL) dosimeters. Measurements were made in air with a standard 16-cm acrylic head phantom and with a standard 32-cm acrylic body phantom. Physical dose measurements determined from the ionization chamber in air for 100 and 120 kVp beam energies were used to derive photon-fluence calibration factors. Both ion chamber and OSL measurement results provide useful comparisons in the validation of the Monte Carlo simulations. It was found that simulated and measured CTDI values were within an overall average of 6% of each other.


Assuntos
Método de Monte Carlo , Imagens de Fantasmas , Polimetil Metacrilato/química , Monitoramento de Radiação , Tomografia Computadorizada por Raios X/métodos , Criança , Simulação por Computador , Humanos , Fótons , Doses de Radiação , Espectrometria gama