1.
IEEE Trans Biomed Eng ; 56(2): 462-70, 2009 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-19342331

RESUMO

A quick wireless label-free detection of disease-related C-reactive proteins (CRPs) using a 200-microm-long microelectromechanical systems (MEMS) microcantilever housed in a 7 x 7 mm(2) reaction chamber with a safe reusable feature is reported. The assay time ranges from about 30 min to 3 h, depending on accuracy. The deflection of the microcantilever due to specific CRP-antiCRP binding is detected using a position-sensitive detector. The converted bio-signal is transmitted by a custom designed wireless amplitude-shift-keying (ASK) transceiver IC fabricated in a 0.18 microm CMOS process. CRP concentrations from 1 microg/mL to 500 microg/mL can be detected. A 0.2-Hz 1-V ac signal instead of traditional bases/acids is applied to the bio-MEMS sensor to unbind the CRP from the microcantilever for reusability.


Assuntos
Técnicas Biossensoriais/instrumentação , Proteína C-Reativa/análise , Sistemas Microeletromecânicos/instrumentação , Nanotecnologia/instrumentação , Processamento de Sinais Assistido por Computador/instrumentação , Doenças Cardiovasculares/diagnóstico , Dimetilpolisiloxanos/química , Desenho de Equipamento , Reutilização de Equipamento , Humanos , Monitorização Ambulatorial/instrumentação