1.
Rev Sci Instrum
; 89(7): 073901, 2018 Jul.
Artigo
em Inglês
| MEDLINE
| ID: mdl-30068119
RESUMO
We describe the development of a broadband (0.3-10 THz) optical pump-terahertz probe spectrometer with an unprecedented combination of temporal resolution (≤200 fs) operating in external magnetic fields as high as 25 T using the new Split Florida-Helix magnet system. Using this new instrument, we measure the transient dynamics in a gallium arsenide four-quantum well sample after photoexcitation at 800 nm.