Serial assessment of coronary artery healing of a biodegradable polymer drug-eluting stent at 1, 2, and 3 months by optical coherence tomography (OCT)­the REPAIR trial
Silva, Guilherme Barreto Gameiro; Meneguz-Moreno, Rafael A; Costa, Ricardo; Chamié, Daniel; Dangas, George; Manica, André; Arruda, Jose Airton; Sousa, Amanda G M R; Feres, Fausto; Costa Jr, J Ribamar.
Affiliation
  • Silva, Guilherme Barreto Gameiro; Department of Interventional Cardiology, Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Meneguz-Moreno, Rafael A; Department of Interventional Cardiology, Instituto Dante Pazzanese de Cardiologia. Department of Medicine, Universidade Federal de Sergipe. São Paulo. BR
  • Costa, Ricardo; Department of Interventional Cardiology, Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Chamié, Daniel; Department of Interventional Cardiology, Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Dangas, George; Department of Interventional Cardiology, Mount Sinai Medical Center. New York. US
  • Manica, André; Instituto de Cardiologia­Fundação Universitária de Cardiologia. Porto Alegre. BR
  • Arruda, Jose Airton; Associação Evangélica Beneficente Espírito Santo. Vitoria. BR
  • Sousa, Amanda G M R; Department of Interventional Cardiology, Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Feres, Fausto; Department of Interventional Cardiology, Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Costa Jr, J Ribamar; Department of Interventional Cardiology, Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
J. invasive cardiol ; 35(5)May. 2023. ilus, graf, tab
Article in En | CONASS, SES-SP, SESSP-IDPCPROD, SES-SP | ID: biblio-1426378
Responsible library: BR79.1
ABSTRACT

BACKGROUND:

Although first-generation drug-eluting stent (DES) devices have effectively achieved their main goal of reducing restenosis, their safety has been limited by suboptimal polymer biocompatibility, delayed stent endothelialization, and local drug toxicity, which ultimately prompted the development of new-generation DES options carrying biocompatible or even biodegradable polymers. Aims. We sought to assess the vessel-healing pattern of the novel sirolimus-eluting Inspiron DES (Scitech Medical) using serial optical coherence tomography (OCT) and assuming the hypothesis that this thin-strut (75-µm), biodegradable-polymer DES promotes a faster healing, with very early strut coverage. METHODS. This is a prospective, multicenter, open-label, single-arm study enrolling 68 patients who underwent percutaneous coronary intervention guided by OCT. These patients were consecutively assigned into 3 groups. The first group had its OCT imaging follow-up performed at 3 months, the second group at 2 months, and the third group at 1 month.

RESULTS:

Mean age was 59.5 years, 70.6% were male, 41.2% had type 2 diabetes, and 29.4% presented with acute coronary syndrome. A total of 72 lesions were treated and 1.06 stents were implanted per patient. OCT assessment of the stents at 1, 2, and 3 months showed a strut coverage of 90.41%, 93.96%, and 97.21%, respectively (P=.04).

CONCLUSION:

The Inspiron DES showed an early strut healing pattern, with >90% of the struts covered by neointima within the first month and with almost all struts covered by the third month.
Subject(s)
Full text: 1 Collection: 06-national / BR Database: CONASS / SES-SP / SESSP-IDPCPROD Main subject: Tomography, Optical Coherence / Acute Coronary Syndrome / Drug-Eluting Stents / Percutaneous Coronary Intervention Type of study: Clinical_trials Language: En Journal: J. invasive cardiol Year: 2023 Document type: Article
Full text: 1 Collection: 06-national / BR Database: CONASS / SES-SP / SESSP-IDPCPROD Main subject: Tomography, Optical Coherence / Acute Coronary Syndrome / Drug-Eluting Stents / Percutaneous Coronary Intervention Type of study: Clinical_trials Language: En Journal: J. invasive cardiol Year: 2023 Document type: Article
...