Your browser doesn't support javascript.
loading
Evaluation of calculation algorithms implemented in different commercial planning systems on an anthropomorphic breast phantom using film dosimetry.
Polednik, Martin; Abo Madyan, Yasser; Schneider, Frank; Wolff, Dirk; Bannach, Burkhardt; Lambrecht, Ulrike; Wallin, André; Cwiekala, Marian; Maurer, Klaus; Reif, Florian; Lohr, Frank; Wenz, Frederik.
Affiliation
  • Polednik M; Department of Radiation Oncology, Mannheim University Hospital, University of Heidelberg, Mannheim Medical Center, Germany. Martin.Polednik@radonk.ma.uni-heidelberg.de
Strahlenther Onkol ; 183(12): 667-72, 2007 Dec.
Article in En | MEDLINE | ID: mdl-18040610
ABSTRACT

PURPOSE:

To evaluate the accuracy of dose calculation algorithms of different planning systems for postoperative tangential radiotherapy in breast cancer. MATERIAL AND

METHODS:

On a CT dataset of an anthropomorphic phantom, a structure set of the left lung, clinical target volume (CTV), planning target volume, heart, and external contour were delineated. The dataset was processed by six radiation oncology centers participating in this multicenter dosimetry project. Conventional plans with two tangential wedged fields were generated in MasterPlan, Pinnacle, Eclipse, TMS, and PrecisePLAN. Plan calculations were done using the beam data of local linacs. The dose distributions were verified under local conditions with Gafchromic-EBT films.

RESULTS:

In all planning systems, deviations between calculation and measurement were around +/-3% in the CTV in the measured plane. Only small areas with deviations of +/-5% were detected. Pencil-beam (PB) calculations overestimated the dose inside the lung by up to 23%. Collapsed cone (CC) underestimated the lung dose by up to 6%.

CONCLUSION:

CC calculates the dose distribution more accurately than PB. Inside regions with electron disequilibrium, however, the dose is slightly underestimated.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Algorithms / Radiotherapy Planning, Computer-Assisted / Software / Breast Neoplasms / Tomography, X-Ray Computed / Anthropometry / Phantoms, Imaging / Film Dosimetry Type of study: Clinical_trials / Diagnostic_studies Limits: Female / Humans Language: En Journal: Strahlenther Onkol Journal subject: NEOPLASIAS / RADIOTERAPIA Year: 2007 Document type: Article Affiliation country:
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Algorithms / Radiotherapy Planning, Computer-Assisted / Software / Breast Neoplasms / Tomography, X-Ray Computed / Anthropometry / Phantoms, Imaging / Film Dosimetry Type of study: Clinical_trials / Diagnostic_studies Limits: Female / Humans Language: En Journal: Strahlenther Onkol Journal subject: NEOPLASIAS / RADIOTERAPIA Year: 2007 Document type: Article Affiliation country:
...