Integration of crawling waves in an ultrasound imaging system. Part 2: signal processing and applications.
Ultrasound Med Biol
; 38(2): 312-23, 2012 Feb.
Article
in En
| MEDLINE
| ID: mdl-22178168
ABSTRACT
This paper introduces methods to generate crawling wave interference patterns from the displacement fields generated from radiation force pushes on a GE Logiq 9 scanner. The same transducer and system provides both the pushing pulses to generate the shear waves and the tracking pulses to measure the displacements. Acoustic power and system limitations result in largely impulsive displacement fields. Measured displacements from pushes on either side of a region-of-interest (ROI) are used to calculate continuously varying interference patterns. This technique is explained along with a brief discussion of the conventional mechanical source-driven crawling waves for comparison. We demonstrate the method on three example cases a gelatin-based phantom with a cylindrical inclusion, an oil-gelatin phantom and mouse livers. The oil-gelatin phantom and the mouse livers demonstrate not only shear speed estimation, but the frequency dependence of the shear wave speeds.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Signal Processing, Computer-Assisted
/
Image Interpretation, Computer-Assisted
/
Computer-Aided Design
/
Elasticity Imaging Techniques