Your browser doesn't support javascript.
loading
Evaluation of a closed-system cytotoxic transfer device in a pharmaceutical isolator.
Vyas, N; Turner, A; Clark, J M; Sewell, G J.
Affiliation
  • Vyas N; Department of Pharmacy, Derriford Hospital, Plymouth, UK School of Geography, Earth and Environmental Sciences, Plymouth University, Plymouth, UK.
  • Turner A; School of Geography, Earth and Environmental Sciences, Plymouth University, Plymouth, UK.
  • Clark JM; Department of Pharmacy, Derriford Hospital, Plymouth, UK.
  • Sewell GJ; School of Health Professions, Plymouth University, Plymouth, UK gjsewell@plymouth.ac.uk.
J Oncol Pharm Pract ; 22(1): 10-9, 2016 Feb.
Article in En | MEDLINE | ID: mdl-25073678
ABSTRACT

PURPOSE:

The occupational risk associated with handling of cytotoxic anticancer drugs is well documented and, in many countries, pharmaceutical isolators are used to contain cytotoxic residues during preparation of cytotoxic infusions. Isolators are difficult to clean leading to concerns that cytotoxic contamination from the work area could be transferred to surfaces of products leaving the isolator. This study investigated the surface contamination arising from the preparation of five anticancer drug infusions (Epirubicin, Fluorouracil, Cisplatin, Oxaliplatin and Carboplatin) in a pharmaceutical isolator and compared use of a conventional syringe and needle technique with a closed-system drug transfer device (CSDTD).

METHODS:

Wipe samples were taken over 1 week from pre-defined areas in the isolator, gloves, preparation mats, and also from the surfaces of prepared cytotoxic infusion bags and pre-filled syringes to obtain baseline surface contamination data. Following operator familiarisation, the CSDTD was then introduced and sampling repeated for a further week (intervention period). The samples obtained were analysed using validated HPLC-UV, HPLC-FL and ICP-MS techniques, as appropriate.

RESULTS:

All surfaces sampled during baseline, including external surfaces of infusions and syringes, were contaminated with each marker drug. During the intervention phase, isolator surfaces were free from detectable contamination and the contamination measured on gloves, preparation mats and surface of infusions was markedly reduced. The frequency of contamination on syringe and infusion surfaces was also lower.

CONCLUSION:

Surface contamination from cytotoxic infusion preparation in a pharmaceutical isolator was significant and could transmit cytotoxic residues to patient and public areas via infusion surfaces. The frequency and amount of contamination were reduced by the CSDTD.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pharmacy / Equipment Contamination / Antineoplastic Agents Language: En Journal: J Oncol Pharm Pract Journal subject: FARMACIA Year: 2016 Document type: Article Affiliation country:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pharmacy / Equipment Contamination / Antineoplastic Agents Language: En Journal: J Oncol Pharm Pract Journal subject: FARMACIA Year: 2016 Document type: Article Affiliation country: