Your browser doesn't support javascript.
loading
Classification of Implantable Rotary Blood Pump States With Class Noise.
IEEE J Biomed Health Inform ; 20(3): 829-837, 2016 05.
Article in En | MEDLINE | ID: mdl-25781963
ABSTRACT
A medical case study related to implantable rotary blood pumps is examined. Five classifiers and two ensemble classifiers are applied to process the signals collected from the pumps for the identification of the aortic valve nonopening pump state. In addition to the noise-free datasets, up to 40% class noise has been added to the signals to evaluate the classification performance when mislabeling is present in the classifier training set. In order to ensure a reliable diagnostic model for the identification of the pump states, classifications performed with and without class noise are evaluated. The multilayer perceptron emerged as the best performing classifier for pump state detection due to its high accuracy as well as robustness against class noise.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Signal Processing, Computer-Assisted / Heart-Assist Devices Limits: Animals Language: En Journal: IEEE J Biomed Health Inform Year: 2016 Document type: Article