  • Szarpak L; From the *Department of Emergency Medicine, Medical University of Warsaw, Warsaw; †Anesthesiology and Intensive Care Unit, Mazovian Regional Hospital, Radom; ‡Department of Nephrological Nursing, Medical University of Warsaw; and §Department of Anesthesiology, Cardinal Wyszynski National Institute of Cardiology, Warsaw, Poland.
Pediatr Emerg Care ; 33(11): 735-739, 2017 Nov.
Article in En | MEDLINE | ID: mdl-27228145
ABSTRACT

OBJECTIVES:

We hypothesized that the Airtraq laryngoscope (Airtraq LLC, Bonita Springs, Fla) is beneficial for intubation of pediatric manikins while performing cardiopulmonary resuscitation (CPR). In the present study, we evaluated the effectiveness of the Macintosh (MAC) laryngoscope (HEINE Optotechnik, Munich, Germany) and Airtraq in 3 simulated CPR scenarios.

METHODS:

A randomized crossover simulation trial was designed. Eighty-three nurses intubated the trachea of a PediaSIM CPR training manikin (FCAE HealthCare, Sarasota, Fla) using the MAC and Airtraq in a normal airway scenario, normal airway with chest compression scenario, and difficult airway with chest compression scenario. The participants were directed to perform a maximum of 3 attempts in each scenario. The success rate, time to intubation, Cormack & Lehane grade, dental compression, and the ease of intubation were measured.

RESULTS:

All participants performed successful intubation with the Airtraq in all 3 scenarios. In all scenarios, the success rate was significantly higher and the time to intubation was significantly shorter with the Airtraq than with the MAC. Glottic visualization using the Cormack-Lehane scale was also better when using Airtraq in all scenarios.

CONCLUSIONS:

In this manikin study, we found that the Airtraq can be used successfully for the intubation of pediatric manikins with normal and difficult airways by medical staff without previous experience in pediatric intubation. Moreover, intubation can be achieved without interrupting chest compression. The use of the Airtraq compared with the MAC led to faster time to intubation. Nevertheless, we recommend that the performance of the Airtraq and the MAC during CPR should be further evaluated in a clinical setting.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Laryngoscopes / Intubation, Intratracheal / Laryngoscopy Type of study: Clinical_trials Limits: Adult / Child / Female / Humans / Male Language: En Journal: Pediatr Emerg Care Journal subject: MEDICINA DE EMERGENCIA / PEDIATRIA Year: 2017 Document type: Article Affiliation country:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Laryngoscopes / Intubation, Intratracheal / Laryngoscopy Type of study: Clinical_trials Limits: Adult / Child / Female / Humans / Male Language: En Journal: Pediatr Emerg Care Journal subject: MEDICINA DE EMERGENCIA / PEDIATRIA Year: 2017 Document type: Article Affiliation country: