The Macek and Davis experiment revisited: a large ring laser interferometer operating on the 2s2 â 2p4 transition of neon.
Appl Opt
; 58(2): 302-307, 2019 Jan 10.
Article
in En
| MEDLINE
| ID: mdl-30645305
ABSTRACT
We operate a large helium-neon-based ring laser interferometer with single-crystal GaAs/AlGaAs optical coatings on the 2s2â2p4 transition of neon at a wavelength of 1.152276 µm. For either single longitudinal- or phase-locked multi-mode operation, the preferable gas composition for gyroscopic operation is 0.2 and 0.3 mbar of 5050 neon with total pressures between 6-12 mbar. The Earth rotation bias is sufficient to unlock the device, yielding a Sagnac frequency of approximately 60 Hz.