Effects of alterations in positron emission tomography imaging parameters on radiomics features.
Ger, Rachel B; Meier, Joseph G; Pahlka, Raymond B; Gay, Skylar; Mumme, Raymond; Fuller, Clifton D; Li, Heng; Howell, Rebecca M; Layman, Rick R; Stafford, R Jason; Zhou, Shouhao; Mawlawi, Osama; Court, Laurence E.
Affiliation
  • Ger RB; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America.
  • Meier JG; MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, United States of America.
  • Pahlka RB; MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, United States of America.
  • Gay S; Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America.
  • Mumme R; Department of Radiology, Texas Children's Hospital, Houston, Texas, United States of America.
  • Fuller CD; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America.
  • Li H; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America.
  • Howell RM; MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, United States of America.
  • Layman RR; Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America.
  • Stafford RJ; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America.
  • Zhou S; MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, United States of America.
  • Mawlawi O; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America.
  • Court LE; MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, United States of America.
PLoS One ; 14(9): e0221877, 2019.
Article in En | MEDLINE | ID: mdl-31487307
ABSTRACT
Radiomics studies require large patient cohorts, which often include patients imaged using different imaging protocols. We aimed to determine the impact of variability in imaging protocol parameters and interscanner variability using a phantom that produced feature values similar to those of patients. Positron emission tomography (PET) scans of a Hoffman brain phantom were acquired on GE Discovery 710, Siemens mCT, and Philips Vereos scanners. A standard-protocol scan was acquired on each machine, and then each parameter that could be changed was altered individually. The phantom was contoured with 10 regions of interest (ROIs). Values for 45 features with 2 different preprocessing techniques were extracted for each image. To determine the impact of each parameter on the reliability of each radiomics feature, the intraclass correlation coefficient (ICC) was calculated with the ROIs as the subjects and the parameter values as the raters. For interscanner comparisons, we compared the standard deviation of each radiomics feature value from the standard-protocol images to the standard deviation of the same radiomics feature from PET scans of 224 patients with non-small cell lung cancer. When the pixel size was resampled prior to feature extraction, all features had good reliability (ICC > 0.75) for the field of view and matrix size. The time per bed position had excellent reliability (ICC > 0.9) on all features. When the filter cutoff was restricted to values below 6 mm, all features had good reliability. Similarly, when subsets and iterations were restricted to reasonable values used in clinics, almost all features had good reliability. The average ratio of the standard deviation of features on the phantom scans to that of the NSCLC patient scans was 0.73 using fixed-bin-width preprocessing and 0.92 using 64-level preprocessing. Most radiomics feature values had at least good reliability when imaging protocol parameters were within clinically used ranges. However, interscanner variability was about equal to interpatient variability; therefore, caution must be used when combining patients scanned on equipment from different vendors in radiomics data sets.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Algorithms / Image Processing, Computer-Assisted / Carcinoma, Non-Small-Cell Lung / Phantoms, Imaging / Positron-Emission Tomography / Lung Neoplasms Type of study: Guideline / Observational_studies / Prognostic_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Journal: PLoS One Journal subject: CIENCIA / MEDICINA Year: 2019 Document type: Article Affiliation country:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Algorithms / Image Processing, Computer-Assisted / Carcinoma, Non-Small-Cell Lung / Phantoms, Imaging / Positron-Emission Tomography / Lung Neoplasms Type of study: Guideline / Observational_studies / Prognostic_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Journal: PLoS One Journal subject: CIENCIA / MEDICINA Year: 2019 Document type: Article Affiliation country: