Gyroscopic performance and some seismic measurements made with a 10 meter perimeter ring laser gyro housed in the Ernest Rutherford building.
Appl Opt ; 60(6): 1737-1743, 2021 Feb 20.
Article in En | MEDLINE | ID: mdl-33690512
ABSTRACT
We describe the construction and operation of a large ring laser whose beam paths enclose an area of 6.25m2. The gyroscopic performance of this large laser interferometer was determined using laser operation at a wavelength of 632.8 nm. The laser cavity Q was inferred to be 1.1×1012 via a measured ring-down time of 375 µs, and the measured Sagnac frequency is 198.40 Hz due to Earth's rotation. The measured experimental sensitivity to rotation achieved is 7.9×10-12rad/s/Hz at an averaging interval of 512 s (being limited primarily by ambient building noise). The observation of microseismic activity in the 200 mHz region as well as local earthquakes is discussed.

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Appl Opt Year: 2021 Document type: Article