Your browser doesn't support javascript.
loading
Safety and efficacy evaluation of the automatic stepwise recruitment maneuver in the neonatal population: An in vivo interventional study. Can anesthesiologists safely perform automatic lung recruitment maneuvers in neonates?
Serrano Zueras, Clara; Guilló Moreno, Verónica; Santos González, Martín; Gómez Nieto, Francisco Javier; Hedenstierna, Göran; García Fernández, Javier.
Affiliation
  • Serrano Zueras C; Department of Anaesthesiology, Intensive Care and Pain, Hospital Universitario Puerta de Hierro en Majadahonda, Majadahonda, Spain.
  • Guilló Moreno V; Department of Anaesthesiology, Intensive Care and Pain, Hospital Universitario Puerta de Hierro en Majadahonda, Majadahonda, Spain.
  • Santos González M; Medical and surgical research unit Instituto de Investigación Sanitaria Puerta de Hierro-Segovia de Arana Hospital Universitario Puerta de Hierro en Majadahonda, Majadahonda, Spain.
  • Gómez Nieto FJ; Department of Anaesthesiology, Intensive Care and Pain, Hospital Universitario Puerta de Hierro en Majadahonda, Majadahonda, Spain.
  • Hedenstierna G; Uppsala University, Uppsala, Sweden.
  • García Fernández J; Department of Anaesthesiology, Intensive Care and Pain, Hospital Universitario Puerta de Hierro en Majadahonda, Majadahonda, Spain.
Paediatr Anaesth ; 31(9): 1003-1010, 2021 Sep.
Article in En | MEDLINE | ID: mdl-34152683
ABSTRACT

BACKGROUND:

A new software has recently been incorporated in almost all new anesthesia machines to enable automatic lung recruitment maneuvers. To date, no studies have assessed the safety and efficacy of these automatic software programs in the neonatal population.

AIMS:

We aimed to evaluate the safety and efficacy of the lung recruitment maneuver performed using the automatic stepwise recruitment maneuver software of the FLOW-i 4.3 Anesthesia System® in a healthy and live neonatal model.

METHODS:

Eight male newborn piglets were included in the study. The lung recruitment maneuver was performed in pressure-controlled ventilation with a constant driving pressure (15 cmH2 O) in a stepwise increasing positive end-expiratory pressure (PEEP) model. The target peak inspiratory pressure (PIP) was 30 cmH2 O and PEEP was 15 cmH2 O. The maneuver lasted for 39 seconds. The hemodynamic variables were monitored using the PICCO® system. The following respiratory parameters were monitored oxygen saturation, fraction of inspired oxygen, partial pressure of oxygen and carbon dioxide in the arterial blood, end-tidal carbon dioxide pressure, PIP, plateau pressure, PEEP, static compliance (Cstat ), and dynamic compliance (Cdyn ). Safety was evaluated by assessing the accuracy of the software, need for not interrupting the maneuver, hemodynamic stability, and absence of adverse respiratory events with the lung recruitment maneuver. Efficacy was evaluated by improvement in Cstat and Cdyn after performing the lung recruitment maneuver.

RESULTS:

All lung recruitment maneuvers were safely performed as scheduled without any interruptions. No pneumothorax or other side effects were observed. Hemodynamic stability was maintained during the lung recruitment maneuver. We observed an improvement of 33% in Cdyn and 24% in Cstat after the maneuver.

CONCLUSIONS:

The automatic stepwise recruitment maneuver software of the FLOW-i 4.3 Anesthesia System® is safe and efficacious in a healthy neonatal model. We did not observe any adverse respiratory or hemodynamic events during the implementation of the lung recruitment maneuver in the pressure-controlled ventilation mode using a stepwise increasing PEEP (30/15 cmH2 O) approach.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Positive-Pressure Respiration / Anesthesiologists Type of study: Prognostic_studies Limits: Animals / Humans / Male Language: En Journal: Paediatr Anaesth Journal subject: ANESTESIOLOGIA / PEDIATRIA Year: 2021 Document type: Article Affiliation country:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Positive-Pressure Respiration / Anesthesiologists Type of study: Prognostic_studies Limits: Animals / Humans / Male Language: En Journal: Paediatr Anaesth Journal subject: ANESTESIOLOGIA / PEDIATRIA Year: 2021 Document type: Article Affiliation country:
...