Estimating the Precision of Quantitative Imaging Biomarkers without Test-Retest Studies.
Obuchowski, Nancy A; Buckler, Andrew J.
Affiliation
  • Obuchowski NA; Quantitative Health Sciences, JJN3, Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195. Electronic address: obuchon@ccf.org.
  • Buckler AJ; Elucid Bioimaging Inc., 2 Park Plaza, Boston, MA 02116.
Acad Radiol ; 29(4): 543-549, 2022 04.
Article in En | MEDLINE | ID: mdl-34272163
ABSTRACT
RATIONALE AND

OBJECTIVES:

A critical performance metric for any quantitative imaging biomarker is its ability to reliably generate similar values on repeat testing. This is known as the repeatability of the biomarker, and it is used to determine the minimum detectable change needed in order to show that a change over time is real change and not just due to measurement error. Test-retest studies are the classic approach for estimating repeatability; however, these studies can be infeasible when the imaging is expensive, time-consuming, invasive, or requires contrast agents. The objective of this study was to develop and test a method for estimating repeatability without a test-retest study. MATERIALS AND

METHODS:

We present a statistical method for estimating repeatability and testing whether an imaging method meets a specified criterion for repeatability in the absence of a test-retest study. The new method is applicable for the particular situation where a reference standard is available. A Monte Carlo simulation study was conducted to evaluate the performance of the new method.

RESULTS:

The proposed estimator is unbiased, and hypothesis tests with the new estimator have nominal type I error rate and power similar to a test-retest study. We considered the situation where the reference standard provides the true value, as well as when the reference standard itself has various magnitudes of measurement error. An example from CT imaging biomarkers of atherosclerosis illustrates the new method.

CONCLUSION:

Precision of a QIB can be measured without a test-retest study in the situation where a reference standard is available.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Diagnostic Imaging / Contrast Media Type of study: Diagnostic_studies / Health_economic_evaluation Limits: Humans Language: En Journal: Acad Radiol Journal subject: RADIOLOGIA Year: 2022 Document type: Article