Your browser doesn't support javascript.
loading
Safety of gadolinium based contrast agents in magnetic resonance imaging-guided radiotherapy - An investigation of chelate stability using relaxometry.
Mahmood, Faisal; Nielsen, Ulla Gro; Jørgensen, Christian Brandt; Brink, Carsten; Thomsen, Henrik S; Hansen, Rasmus Hvass.
Affiliation
  • Mahmood F; Laboratory of Radiation Physics, Department of Oncology, Odense University Hospital, Kloevervaenget 19, Indgang 85, Pavillonen, Stuen, 5000 Odense C, Denmark.
  • Nielsen UG; Department of Clinical Research, University of Southern Denmark, Winsløwparken 19, 3. Sal, Odense C 5000, Denmark.
  • Jørgensen CB; Department of Physics, Chemistry and Pharmacy, University of Southern Denmark, Campusvej 55, 5230 Odense M, Denmark.
  • Brink C; Department of Physics, Chemistry and Pharmacy, University of Southern Denmark, Campusvej 55, 5230 Odense M, Denmark.
  • Thomsen HS; Laboratory of Radiation Physics, Department of Oncology, Odense University Hospital, Kloevervaenget 19, Indgang 85, Pavillonen, Stuen, 5000 Odense C, Denmark.
  • Hansen RH; Department of Clinical Research, University of Southern Denmark, Winsløwparken 19, 3. Sal, Odense C 5000, Denmark.
Phys Imaging Radiat Oncol ; 21: 96-100, 2022 Jan.
Article in En | MEDLINE | ID: mdl-35243039
ABSTRACT
BACKGROUND AND

PURPOSE:

With the introduction of hybrid magnetic resonance linacs (MR-linac), improved imaging has enabled daily treatment adaptation. However, the use of gadolinium based contrast agents (GBCAs) is desired to further improve MR image contrast. GBCAs are in the form of a non-toxic metalorganic gadolinium complex, but toxic un-chelated aqueous gadolinium(III), Gd3+(aq), can be released in patients if the organic ligand is degraded by the radiation. In this study, T1 relaxation measurements were performed to study the effect of radiation on three GBCAs. MATERIALS AND

METHODS:

GBCAs, gadoteric acid, gadobutrol and gadoxectic acid were investigated in a concentration range of 10-100 mM. Measurements were performed on a 500 MHz nuclear MR (NMR) spectrometer with a high-resolution inversion recovery sequence to determine T1. Samples were irradiated with 7 MV photons on an MR-linac to a total dose of 100 Gy. The lower detection limit of Gd3+(aq) was established by estimating the overall measurement uncertainty and comparing to corresponding changes in R1 when replacing chelated Gd3+ with gadolinium nitrate at predefined percentages.

RESULTS:

The overall measurement uncertainty was estimated to ±0.0053 ms-1, corresponding to Gd3+(aq) detection levels 1%-1.5% or 1-4.5 micro molar at clinical GBCA dosage. No detectable differences in R1 were observed between irradiated and non-irradiated samples for any GBCA.

CONCLUSIONS:

This study did not find any measurable degradation of GBCAs due to irradiation with high-energy X-rays, however, in-vivo investigations are needed to provide the clinical basis for safe use of contrast agents in a radiotherapy workflow.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Phys Imaging Radiat Oncol Year: 2022 Document type: Article Affiliation country:

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Phys Imaging Radiat Oncol Year: 2022 Document type: Article Affiliation country:
...