Validation of the STERIS Amsco 630LS steam sterilizer autoclave for inactivation of category a medical waste from patients with high-consequence infectious diseases.
Tarabay, Jessica A; Morgan, Jill S; Kraft, Colleen S; Sexton, Mary Elizabeth.
Affiliation
  • Tarabay JA; Emory Healthcare, Atlanta, Georgia.
  • Morgan JS; Emory Healthcare, Atlanta, Georgia.
  • Kraft CS; Division of Infectious Diseases, Emory University School of Medicine, Atlanta, Georgia.
  • Sexton ME; Department of Pathology and Laboratory Medicine Emory University School of Medicine, Atlanta, Georgia.
Infect Control Hosp Epidemiol ; 45(2): 260-262, 2024 Feb.
Article in En | MEDLINE | ID: mdl-37675506
ABSTRACT
Hospitals caring for patients with high-consequence pathogens may need to safely manage large volumes of category A waste. Using biological indicators to assess for successful sterilization, autoclave cycle parameters that would inactivate 4 categories of waste were identified and validated utilizing a STERIS Amsco 630LS Steam Sterilizer.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Communicable Diseases / Medical Waste Limits: Humans Language: En Journal: Infect Control Hosp Epidemiol Journal subject: DOENCAS TRANSMISSIVEIS / ENFERMAGEM / EPIDEMIOLOGIA / HOSPITAIS Year: 2024 Document type: Article Affiliation country:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Communicable Diseases / Medical Waste Limits: Humans Language: En Journal: Infect Control Hosp Epidemiol Journal subject: DOENCAS TRANSMISSIVEIS / ENFERMAGEM / EPIDEMIOLOGIA / HOSPITAIS Year: 2024 Document type: Article Affiliation country:
...