ABSTRACT
PURPOSE:
Liver cirrhosis disrupts liver function and tissue perfusion, detectable by magnetic resonance imaging (MRI). Assessing liver function at the voxel level with 13-b value intravoxel incoherent motion diffusion-weighted imaging (IVIM-DWI) could aid in radiation therapy liver-sparing treatment for patients with early impairment. This study aimed to evaluate the feasibility of IVIM-DWI for liver function assessment and correlate it with other multiparametric (mp) MRI methods at the voxel level.METHOD:
This study investigates the variability of apparent diffusion coefficient (ADC) derived from 13-b value IVIM-DWI and B1-corrected dual flip angle (DFA) T1 mapping. Experiments were conducted in-vitro with QIBA and NIST phantoms and in 10 healthy volunteers for IVIM-DWI. Additionally, 12 patients underwent an mp-MRI examination. The imaging protocol included a 13-b value IVIM-DWI sequence for generating IVIM parametric maps. B1-corrected DFA T1 pulse sequence was used for generating T1 maps, and Gadoxatate low temporal resolution dynamic contrast-enhanced (LTR-DCE) MRI was used for generating the Hepatic extraction fraction (HEF) map. The Mann-Whitney U test was employed to compare IVIM-DWI parameters (Pure Diffusion, Dslow ; Pseudo diffusion, Dfast ; and Perfusion Fraction, Fp ) between the healthy volunteer and patient groups. Furthermore, in the patient group, statistical correlations were assessed at a voxel level between LTR-DCE MRI-derived HEF, T1 post-Gadoxetate administration, ΔT1%, and various IVIM parameters using Pearson correlation.RESULTS:
For-vitro measurements, the maximum coefficient of variation of the ADC and T1 parameters was 12.4% and 16.1%, respectively. The results also showed that Fp and Dfast were able to distinguish between healthy liver function and mild liver function impairment at the global level, with p = 0.002 for Fp and p < 0.001 for Dfast . Within the patient group, these parameters also exhibited a moderate correlation with HEF at the voxel level.CONCLUSION:
Overall, the study highlighted the potential of Dfast and Fp for detecting liver function impairment at both global and pixel levels.Key words