Use of volumetric absorptive microsampling and parallel reaction monitoring mass spectrometry for tacrolimus blood trough measurements at home in pediatric heart transplant patients.
Zhao, Junfang; Setchell, Kenneth D R; Zhao, Xueheng; Galandi, Stephanie; Garr, BreAnn N; Gao, Zhiqian; Chin, Clifford; Stark, Shelly; Steele, Paul E; Ryan, Thomas D.
Affiliation
  • Zhao J; Division of Pathology & Laboratory Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Setchell KDR; Division of Pathology & Laboratory Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Zhao X; Department of Pediatrics, University of Cincinnati College of Medicine, Cincinnati, OH, USA.
  • Galandi S; Division of Pathology & Laboratory Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Garr BN; Department of Pediatrics, University of Cincinnati College of Medicine, Cincinnati, OH, USA.
  • Gao Z; Division of Pathology & Laboratory Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Chin C; Heart Institute, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Stark S; Heart Institute, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Steele PE; Heart Institute, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Ryan TD; Department of Pediatrics, University of Cincinnati College of Medicine, Cincinnati, OH, USA.
J Mass Spectrom Adv Clin Lab ; 31: 1-7, 2024 Jan.
Article in En | MEDLINE | ID: mdl-38163003
ABSTRACT

Background:

Measurement of trough levels for calcineurin inhibitors by venipuncture sampling is a mainstay of patient management in solid organ transplant recipients but challenging in pediatric patients. Volumetric Absorptive Microsampling (VAMS) is a patient-friendly, minimally invasive sampling technique to accurately collect blood. An assay for measurement of tacrolimus in blood using VAMS, coupled with parallel reaction monitoring (PRM) mass spectrometry, was validated in pediatric heart transplant patients.

Methods:

Tacrolimus was measured by a newly developed high-resolution PRM assay and compared with low-resolution tandem mass spectrometry (MRM). Dried blood samples were collected from pediatric heart transplant patients (n = 35) using VAMS devices and a satisfaction survey was completed by patients/guardians. Tacrolimus concentrations were compared across whole liquid blood, dried blood spots, and capillary blood, and shipping stability determined.

Results:

The PRM assay was linear over a range 1-50 ng/mL, similar to MRM but had greater specificity due to reduced background noise. No significant differences in tacrolimus concentrations were observed between VAMS and venous blood. Tacrolimus dried on VAM tips was stable for 14 days and concentrations were unaffected by postal shipping. The variability in two simultaneously collected at-home patient samples was minimal - average concentration difference was 0.12 ± 0.94 ng/mL (p = 0.6) between paired samples.

Conclusion:

A high resolution PRM mass spectrometry assay was developed for home-based dried blood collections for therapeutic monitoring of tacrolimus. The advantage of PRM was enhanced specificity and the VAMS devices provided a simple and convenient approach to blood sampling at home in pediatric heart transplant patients.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Mass Spectrom Adv Clin Lab Year: 2024 Document type: Article Affiliation country: Country of publication:

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Mass Spectrom Adv Clin Lab Year: 2024 Document type: Article Affiliation country: Country of publication: