Setup and characterisation according to NEMA NU 4 of thephenoPET scanner, a PET system dedicated for plant sciences.
Phys Med Biol
; 69(5)2024 Feb 23.
Article
in En
| MEDLINE
| ID: mdl-38271724
ABSTRACT
Objective.ThephenoPET system is a plant dedicated positron emission tomography (PET) scanner consisting of fully digital photo multipliers with lutetium-yttrium oxyorthosilicate crystals and located inside a custom climate chamber. Here, we present the setup ofphenoPET, its data processing and image reconstruction together with its performance.Approach.The performance characterization follows the national electrical manufacturers association (NEMA) standard for small animal PET systems with a number of adoptions due to the vertical oriented bore of a PET for plant sciences. In addition temperature stability and spatial resolution with a hot rod phantom are addressed.Main results.The spatial resolution for a22Na point source at a radial distance of 5 mm to the center of the field-of-view (FOV) is 1.45 mm, 0.82 mm and 1.88 mm with filtered back projection in radial, tangential and axial direction, respectively. A hot rod phantom with18F gives a spatial resolution of up to 1.6 mm. The peak noise-equivalent count rates are 550 kcps @ 35.08 MBq, 308 kcps @ 33 MBq and 45 kcps @ 40.60 MBq for the mouse, rat and monkey size scatter phantoms, respectively. The scatter fractions for these phantoms are 12.63%, 22.64% and 55.90%. We observe a peak sensitivity of up to 3.6% and a total sensitivity of up toSA,tot= 2.17%. For the NEMA image quality phantom we observe a uniformity of %STD= 4.22% with ordinary Poisson maximum likelihood expectation-maximization with 52 iterations. Here, recovery coefficients of 0.12, 0.64, 0.89, 0.93 and 0.91 for 1 mm, 2 mm, 3 mm, 4 mm and 5 mm rods are obtained and spill-over ratios of 0.08 and 0.14 for the water-filled and air-filled inserts, respectively.Significance.ThephenoPET and its laboratory are now in routine operation for the administration of [11C]CO2and non-invasive measurement of transport and allocation of11C-labelled photoassimilates in plants.
Key words
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Image Processing, Computer-Assisted
/
Positron-Emission Tomography
Limits:
Animals
Language:
En
Journal:
Phys Med Biol
Year:
2024
Document type:
Article
Affiliation country: