Your browser doesn't support javascript.
loading
DCS blood flow index underestimates skeletal muscle perfusion in vivo: rationale and early evidence for the NIRS-DCS perfusion index.
Bartlett, Miles F; Oneglia, Andrew P; Ricard, Mark D; Siddiqui, Ashfaq; Englund, Erin K; Buckley, Erin M; Hueber, Dennis M; Nelson, Michael D.
Affiliation
  • Bartlett MF; University of Texas at Arlington, Arlington, Texas, United States.
  • Oneglia AP; Bartlett Sciences LLC, Dallas, Texas, United States.
  • Ricard MD; University of Texas at Arlington, Arlington, Texas, United States.
  • Siddiqui A; University of Texas at Arlington, Arlington, Texas, United States.
  • Englund EK; Total Vascular Wellness, Mansfield, Texas, United States.
  • Buckley EM; University of Colorado Anschutz Medical Campus, Aurora, Colorado, United States.
  • Hueber DM; Georgia Institute of Technology and Emory University, Atlanta, Georgia, United States.
  • Nelson MD; ISS Inc., Champaign, Illinois, United States.
J Biomed Opt ; 29(2): 020501, 2024 Feb.
Article in En | MEDLINE | ID: mdl-38322728
ABSTRACT

Significance:

Diffuse correlation spectroscopy (DCS) permits non-invasive assessment of skeletal muscle blood flow but may misestimate changes in muscle perfusion.

Aim:

We aimed to highlight recent evidence that DCS blood flow index (BFI) misestimates changes in muscle blood flow during physiological perturbation and to introduce a novel approach that adjusts BFI for estimated changes in vasodilation.

Approach:

We measured changes in muscle BFI during quadriceps and forearm exercises using DCS, the latter of which were adjusted for estimated changes in microvascular flow area and then compared to Doppler ultrasound in the brachial artery. Then, we compared adjusted BFI- and arterial spin labeling (ASL) MRI measures of gastrocnemius blood flow during reactive hyperemia and plantar flexion exercise.

Results:

We observed little-to-no change in quadriceps BFI during maximal-effort exercise. Similarly, forearm BFI was modestly increased during handgrip exercise, but the magnitude was significantly lower than measured by Doppler ultrasound in the brachial artery. However, this difference was ameliorated after adjusting BFI for estimated changes in microvascular flow area. Similar observations were also observed in the gastrocnemius when directly comparing the adjusted BFI values to ASL-MRI.

Conclusions:

Adjusting BFI for estimated changes in microvascular flow area may improve DCS estimates of muscle blood flow, but further study is needed to validate these methods moving forward.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Hand Strength / Perfusion Index Language: En Journal: J Biomed Opt / J. biomed. opt. (Online) / Journal of biomedical optics (Online) Journal subject: ENGENHARIA BIOMEDICA / OFTALMOLOGIA Year: 2024 Document type: Article Affiliation country: Country of publication:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Hand Strength / Perfusion Index Language: En Journal: J Biomed Opt / J. biomed. opt. (Online) / Journal of biomedical optics (Online) Journal subject: ENGENHARIA BIOMEDICA / OFTALMOLOGIA Year: 2024 Document type: Article Affiliation country: Country of publication: