ST-Segment Elevation During Percutaneous Left Atrial Appendage Closure.
JACC Case Rep
; 29(4): 102216, 2024 Feb 21.
Article
in En
| MEDLINE
| ID: mdl-38379648
ABSTRACT
A 70-year-old patient with paroxysmal atrial fibrillation underwent left atrial appendage closure. The patient experienced transient hypotension during device implantation. The procedure was abandoned because of ST-T-wave changes on electrocardiography and elevated coronary flow velocity on transesophageal echocardiography, which indicated that the device caused coronary artery compression.