[Accuracy Verification for Slice Sensitivity Profile Measurement Method by Averaging the Multiple Tilted Wires in Computed Tomography Image].
Fukaya, Takeshi; Murakami, Katsuhiko; Saito, Masaki; Katakura, Toshihiko.
Affiliation
  • Fukaya T; Department of Radiology, Public University Corporation Fukushima Medical University Hospital.
  • Murakami K; Department of Radiology, Public University Corporation Fukushima Medical University Hospital.
  • Saito M; Department of Radiology, Public University Corporation Fukushima Medical University Hospital.
  • Katakura T; B.B.M.I.L.
Nihon Hoshasen Gijutsu Gakkai Zasshi ; 80(4): 390-402, 2024 Apr 20.
Article in Ja | MEDLINE | ID: mdl-38447996
ABSTRACT

PURPOSE:

Several studies present the unsuitability of the tilted-wire method for slice sensitivity profile (SSP) in helical scan. We compared the accuracy for SSP by the tilted-wire averaging method using multiple wire profiles and by the conventional micro-coin method.

METHODS:

A micro-coin phantom positioned at the center or the off-center was scanned using a 64-detector row CT scanner in different positions where an X-ray tube starts scanning. In the same way, tilted-wire averaging phantoms, approximately 70 mm in diameter, in the shape of a donut, 8 wires tilted from the circumference toward the center, were scanned. Images were reconstructed with a slice thickness of 0.5 mm.

RESULTS:

The relative errors of full width at half maximum (FWHM) by the tilted-wire averaging method were -0.015 mm to -0.004 mm (-1.98% to -0.56%) at the center compared to those by the micro-coin method, and it is almost the same value regardless of the number of wires. Relative errors were 0.001 mm to 0.029 mm (0.11% to 3.74%) at the upper 8 cm from the center, and 0.014 mm to 0.078 mm (1.86% to 10.25%) at the upper 16 cm, and the value of relative errors increased as it got farther from the center and as the number of wires went fewer.

CONCLUSION:

This study indicated that accurate measurement of SSP may be achieved by using 4 (arranged every 90 degrees) or more averaging wires.