Your browser doesn't support javascript.
loading
On the design of a macro-micro parallel manipulator for cochlear microrobot operations.
Gezgin, Erkin; Yasar, Alp E; Uslu, Tugrul; Koçak, Mertcan; Güzin, Didem; Özbek, Seda; Türkmen, Gökmen A; Karayaman, Goncagül; Alsangur, Rahime; Bidikli, Baris; Can, Fatih C; Çetin, Levent.
Affiliation
  • Gezgin E; Department of Mechatronics Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Yasar AE; Department of Mechanical Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Uslu T; Department of Mechanical Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Koçak M; Department of Mechatronics Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Güzin D; Department of Mechanical Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Özbek S; Department of Mechanical Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Türkmen GA; Department of Mechanical Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Karayaman G; Department of Mechatronics Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Alsangur R; Department of Mechatronics Engineering, Dokuz Eylül University, Alsancak, Turkey.
  • Bidikli B; Department of Mechatronics Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Can FC; Department of Mechatronics Engineering, Izmir Katip Celebi University, Izmir, Turkey.
  • Çetin L; Department of Mechatronics Engineering, Izmir Katip Celebi University, Izmir, Turkey.
Int J Med Robot ; 20(4): e2654, 2024 Aug.
Article in En | MEDLINE | ID: mdl-38941214
ABSTRACT

BACKGROUND:

The method of stem cell transfer to narrow cochlear canals in vivo to generate hair cells is still an unclear operation. Thus, the development of any possible method that will ensure the usage of medical microrobots in small cochlear workspaces is a challenging procedure.

METHODS:

The current study tries to introduce a macro-micro manipulator system composed of a 6-DoF industrial serial manipulator as a macro manipulator and a proposed 5-DoF parallel manipulator with dual end effectors as a micro manipulator carrying permanent magnets for tetherless microrobot actuation inside the cochlea.

RESULTS:

Throughout the study, structural synthesis and kinematic analysis of the proposed micro manipulator were introduced. A prototype of the manipulator was manufactured and its hardware verification procedures were carried out using motion capture cameras and surgical navigation registration methodologies.

CONCLUSIONS:

Following motion training, the assembled macro-micro manipulator was successfully utilised to actuate a microrobot placed inside a manufactured cochlea mockup model.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Cochlea / Equipment Design / Robotic Surgical Procedures Limits: Humans Language: En Journal: Int J Med Robot Year: 2024 Document type: Article Affiliation country:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Cochlea / Equipment Design / Robotic Surgical Procedures Limits: Humans Language: En Journal: Int J Med Robot Year: 2024 Document type: Article Affiliation country: