Pilot study on optimizing pressure for standardized capillary refill time measurement.
Ma, Zi-Yu; Sun, Shen; Wu, Shui-Cai; Lin, Lan; Chen, Yi-Xiong; Zhao, Dong; Morgan, Stephen P.
Affiliation
  • Ma ZY; Faculty of Environment and Life, Beijing University of Technology, China, No. 100 Pingleyuan, Chaoyang District, Beijing, 100124, China.
  • Sun S; Faculty of Environment and Life, Beijing University of Technology, China, No. 100 Pingleyuan, Chaoyang District, Beijing, 100124, China.
  • Wu SC; Faculty of Environment and Life, Beijing University of Technology, China, No. 100 Pingleyuan, Chaoyang District, Beijing, 100124, China.
  • Lin L; Faculty of Environment and Life, Beijing University of Technology, China, No. 100 Pingleyuan, Chaoyang District, Beijing, 100124, China.
  • Chen YX; Beijing Science and Technology Project Manager Management Corporation Ltd, Beijing, China.
  • Zhao D; Department of Endocrinology, Beijing LuHe Hospital, Capital Medical University, Beijing, China.
  • Morgan SP; Optics and Photonics Research Group, University of Nottingham, Nottingham, UK.
Heliyon ; 10(15): e35716, 2024 Aug 15.
Article in En | MEDLINE | ID: mdl-39170323
ABSTRACT

Purpose:

Capillary Refill Time (CRT) measurement has gained increasing attention in the field of sepsis and septic shock. Recognizing pressure as a fundamental determinant in CRT measurement is crucial for establishing a standardized CRT measurement procedure. In this preliminary study, we elucidated the optimal pressing strength for CRT measurement by analyzing the CRTs measured under varying pressures.

Method:

Seventeen healthy individuals were enlisted to undergo CRT tests on their fingertips at various pressure levels. The applied force was initiated at 0.5N and incrementally increased by 0.5N until it reached 10.5N. An integrated Photoplethysmography (PPG) device was employed to capture fluctuations in light intensity. The CRT was automatically derived from the PPG signals via a specialized algorithm. The study included correlation assessment and reliability evaluation. Box plot and Bland-Altman plot were used to visualize the impact of pressure levels on CRTs.

Results:

A dataset of 1414 CRTs across 21 pressures showed significant differences (Kruskal-Wallis test, p < 0.0001), highlighting the impact of pressure on CRT. CRT values between 4.5N and 10.5N pressures varied less, with an Intraclass Correlation Coefficient (ICC) of 0.499 indicating moderate consistency. Notably, CRTs at 10N and 10.5N pressures revealed a high ICC of 0.790, suggesting strong agreement.

Conclusion:

A pressure range of 4.5N-10.5N is recommended for stable CRT measurements, with 10.0N-10.5N providing optimal consistency and reliability.
Key words