A new method to detect systemic inflammatory response syndrome by continuous monitoring system: an experimental study in rats.
J Med Eng Technol
; 29(1): 18-21, 2005.
Article
in En
| MEDLINE
| ID: mdl-15764377
ABSTRACT
A continuous monitoring system was designed for detection of systemic inflammatory response syndrome (SIRS) using spectrum analyser and local temperature mapping system with deep body thermometer. This method demonstrated significant increases of heart and respiratory rates and deep body temperature from normal levels within 120 minutes after lipopolysaccharide administration in rats. Our monitoring system appears promising for clinical detection of SIRS in the early stages to prevent multiple organ failure.
Search on Google
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Spectrum Analysis
/
Thermography
/
Diagnosis, Computer-Assisted
/
Systemic Inflammatory Response Syndrome
/
Electrocardiography
/
Monitoring, Physiologic
Type of study:
Diagnostic_studies
/
Evaluation_studies
Limits:
Animals
Language:
En
Journal:
J Med Eng Technol
Year:
2005
Document type:
Article
Affiliation country:
Japón