A 9 keV electron-impact liquid-gallium-jet x-ray source.
Otendal, M; Tuohimaa, T; Vogt, U; Hertz, H M.
Affiliation
  • Otendal M; Biomedical and X-Ray Physics, Department of Applied Physics, Royal Institute of Technology/Albanova, SE-10691 Stockholm, Sweden. mikael.otenda@biox.kth.se
Rev Sci Instrum ; 79(1): 016102, 2008 Jan.
Article in En | MEDLINE | ID: mdl-18248074
ABSTRACT
We demonstrate a high-brightness compact 9 keV electron-impact microfocus x-ray source based on a liquid-gallium-jet anode. A approximately 30 W, 50 kV electron gun is focused onto the approximately 20 ms, 30 mum diameter liquid-gallium-jet anode to produce an approximately 10 microm full width at half maximum x-ray spot. The peak spectral brightness is >2 x 10(10) photons(s mm(2) mrad(2)x0.1% BW). Calculation and experiments show potential for increasing this brightness by approximately three orders of magnitude, making the source suitable for laboratory-scale x-ray crystallography and hard x-ray microscopy.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Synchrotrons / Electrons / Gallium Type of study: Diagnostic_studies Language: En Journal: Rev Sci Instrum Year: 2008 Document type: Article Affiliation country: Suecia