The effect of the dc bias voltage on the x-ray bremsstrahlung and beam intensities of medium and highly charged ions of argon.
Rodrigues, G; Lakshmy, P S; Baskaran, R; Kanjilal, D; Roy, A.
Affiliation
  • Rodrigues G; Inter University Accelerator Centre, Aruna Asaf Ali Marg, New Delhi 110067, India. gerosro@gmail.com
Rev Sci Instrum ; 81(2): 02A323, 2010 Feb.
Article in En | MEDLINE | ID: mdl-20192344
ABSTRACT
X-ray bremsstrahlung measurements from the 18 GHz High Temperature Superconducting Electron Cyclotron Resonance Ion Source, Pantechnik-Delhi Ion Source were measured as a function of negative dc bias voltage, keeping all other source operating parameters fixed and the extraction voltage in the off condition. The optimization of medium and highly charged ions of argon with similar source operating parameters is described. It is observed that the high temperature component of the electron is altered significantly with the help of bias voltage, and the electron population has to be maximized for obtaining higher current.

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Rev Sci Instrum Year: 2010 Document type: Article Affiliation country: India