Improved x-ray spectroscopy with room temperature CZT detectors.
Fritz, Shannon G; Shikhaliev, Polad M; Matthews, Kenneth L.
Affiliation
  • Fritz SG; Imaging Physics Laboratory, Department of Physics and Astronomy, Louisiana State University, Baton Rouge, LA 70803, USA.
Phys Med Biol ; 56(17): 5735-51, 2011 Sep 07.
Article in En | MEDLINE | ID: mdl-21841213
ABSTRACT
Compact, room temperature x-ray spectroscopy detectors are of interest in many areas including diagnostic x-ray imaging, radiation protection and dosimetry. Room temperature cadmium zinc telluride (CZT) semiconductor detectors are promising candidates for these applications. One of the major problems for CZT detectors is low-energy tailing of the energy spectrum due to hole trapping. Spectral post-correction methods to correct the tailing effect do not work well for a number of reasons; thus it is advisable to eliminate the hole trapping effect in CZT using physical methods rather than correcting an already deteriorated energy spectrum. One method is using a CZT detector with an electrode configuration which modifies the electric field in the CZT volume to decrease low-energy tailing. Another method is to irradiate the CZT surface at a tilted angle, which modifies depth of interaction to decrease low-energy tailing. Neither method alone, however, eliminates the tailing effect. In this work, we have investigated the combination of modified electric field and tilted angle irradiation in a single detector to further decrease spectral tailing. A planar CZT detector with 10 × 10 × 3 mm³ size and CZT detector with 5 × 5 × 5 mm³ size and cap-shaped electrode were used in this study. The cap-shaped electrode (referred to as CAPture technology) modifies the electric field distribution in the CZT volume and decreases the spectral tailing effect. The detectors were investigated at 90° (normal) and 30° (tilted angle) irradiation modes. Two isotope sources with 59.6 and 122 keV photon energies were used for gamma-ray spectroscopy experiments. X-ray spectroscopy was performed using collimated beams at 60, 80 and 120 kVp tube voltages, in both normal and tilted angle irradiation. Measured x-ray spectra were corrected for K x-ray escape fractions that were calculated using Monte Carlo methods. The x-ray spectra measured with tilted angle CAPture detector at 60, 80 and 120 kVp tube voltages were compared to corresponding theoretical spectra. The low-energy tailing was nearly completely eliminated from 59.6 and 122 keV isotope spectra, and 60, 80 and 120 kVp x-ray spectra, when CAPture detector was used with 30° tilted angle irradiation. It is concluded that using a CZT detector with modified electric field in tilted angle configuration resolves problem of the tailing effect in CZT detectors, opening promising possibilities in gamma-ray and x-ray spectroscopy applications.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Radiometry / Semiconductors / Spectrometry, X-Ray Emission / Tellurium / Zinc / Cadmium / Monte Carlo Method Type of study: Diagnostic_studies / Health_economic_evaluation Language: En Journal: Phys Med Biol Year: 2011 Document type: Article Affiliation country: Estados Unidos

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Radiometry / Semiconductors / Spectrometry, X-Ray Emission / Tellurium / Zinc / Cadmium / Monte Carlo Method Type of study: Diagnostic_studies / Health_economic_evaluation Language: En Journal: Phys Med Biol Year: 2011 Document type: Article Affiliation country: Estados Unidos