Your browser doesn't support javascript.
loading
Volumetric feedback ablation of uterine fibroids using magnetic resonance-guided high intensity focused ultrasound therapy.
Voogt, M J; Trillaud, H; Kim, Y S; Mali, W P Th M; Barkhausen, J; Bartels, L W; Deckers, R; Frulio, N; Rhim, H; Lim, H K; Eckey, T; Nieminen, H J; Mougenot, C; Keserci, B; Soini, J; Vaara, T; Köhler, M O; Sokka, S; van den Bosch, Maurice A A J.
Affiliation
  • Voogt MJ; Department of Radiology, University Medical Center Utrecht, Heidelberglaan 100, 3584 GX, Utrecht, The Netherlands.
Eur Radiol ; 22(2): 411-7, 2012 Feb.
Article in En | MEDLINE | ID: mdl-21901565
ABSTRACT

OBJECTIVE:

The purpose of this prospective multicenter study was to assess the safety and technical feasibility of volumetric Magnetic Resonance-guided High Intensity Focused Ultrasound (MR-HIFU) ablation for treatment of patients with symptomatic uterine fibroids.

METHODS:

Thirty-three patients with 36 fibroids were treated with volumetric MR-HIFU ablation. Treatment capability and technical feasibility were assessed by comparison of the Non-Perfused Volumes (NPVs) with MR thermal dose predicted treatment volumes. Safety was determined by evaluation of complications or adverse events and unintended lesions. Secondary endpoints were pain and discomfort scores, recovery time and length of hospital stay.

RESULTS:

The mean NPV calculated as a percentage of the total fibroid volume was 21.7%. Correlation between the predicted treatment volumes and NPVs was found to be very strong, with a correlation coefficient r of 0.87. All patients tolerated the treatment well and were treated on an outpatient basis. No serious adverse events were reported and recovery time to normal activities was 2.3 ± 1.8 days.

CONCLUSION:

This prospective multicenter study proved that volumetric MR-HIFU is safe and technically feasible for the treatment of symptomatic uterine fibroids. KEY POINTS • Magnetic-resonance-guided high intensity focused ultrasound allows non-invasive treatment of uterine fibroids. • Volumetric feedback ablation is a novel technology that allows larger treatment volumes • MR-guided ultrasound ablation of uterine fibroids appears safe using volumetric feedback.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Ultrasonic Therapy / Uterine Neoplasms / Magnetic Resonance Imaging / Magnetic Resonance Imaging, Interventional / High-Intensity Focused Ultrasound Ablation / Leiomyoma Type of study: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limits: Adolescent / Adult / Female / Humans / Middle aged Country/Region as subject: Europa Language: En Journal: Eur Radiol Journal subject: RADIOLOGIA Year: 2012 Document type: Article Affiliation country: Países Bajos

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Ultrasonic Therapy / Uterine Neoplasms / Magnetic Resonance Imaging / Magnetic Resonance Imaging, Interventional / High-Intensity Focused Ultrasound Ablation / Leiomyoma Type of study: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limits: Adolescent / Adult / Female / Humans / Middle aged Country/Region as subject: Europa Language: En Journal: Eur Radiol Journal subject: RADIOLOGIA Year: 2012 Document type: Article Affiliation country: Países Bajos