Pretreatment of urine samples with SDS improves direct identification of urinary tract pathogens with matrix-assisted laser desorption ionization-time of flight mass spectrometry.
J Clin Microbiol
; 52(1): 335-8, 2014 Jan.
Article
in En
| MEDLINE
| ID: mdl-24226916
ABSTRACT
We pretreated with SDS 71 urine samples with bacterial counts of >10(5) CFU/ml and matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) identification scores of <2, in order to minimize failure rates. Identification improved in 46.5% of samples, remained unchanged in 49.3%, and worsened in 4.2%. The improvement was more evident for Gram-negative (54.3%) than for Gram-positive (32%) bacteria.