A Multi-Pumping Flow System for In Situ Measurements of Dissolved Manganese in Aquatic Systems.
Sensors (Basel)
; 16(12)2016 Nov 30.
Article
in En
| MEDLINE
| ID: mdl-27916898
A METals In Situ analyzer (METIS) has been used to determine dissolved manganese (II) concentrations in the subhalocline waters of the Gotland Deep (central Baltic Sea). High-resolution in situ measurements of total dissolved Mn were obtained in near real-time by spectrophotometry using 1-(2-pyridylazo)-2-naphthol (PAN). PAN is a complexing agent of dissolved Mn and forms a wine-red complex with a maximum absorbance at a wavelength of 562 nm. Results are presented together with ancillary temperature, salinity, and dissolved O 2 data. Lab calibration of the analyzer was performed in a pressure testing tank. A detection limit of 77 nM was obtained. For validation purposes, discrete water samples were taken by using a pump-CTD system. Dissolved Mn in these samples was determined by an independent laboratory based method (inductively coupled plasma-optical emission spectrometry, ICP-OES). Mn measurements from both METIS and ICP-OES analysis were in good agreement. The results showed that the in situ analysis of dissolved Mn is a powerful technique reducing dependencies on heavy and expensive equipment (pump-CTD system, ICP-OES) and is also cost and time effective.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Language:
En
Journal:
Sensors (Basel)
Year:
2016
Document type:
Article
Affiliation country:
Alemania
Country of publication:
Suiza