Percutaneous Spinal Ablation in a Sheep Model: Protective Capacity of an Intact Cortex, Correlation of Ablation Parameters with Ablation Zone Size, and Correlation of Postablation MRI and Pathologic Findings.
Wallace, A N; Hillen, T J; Friedman, M V; Zohny, Z S; Stephens, B H; Greco, S C; Talcott, M R; Jennings, J W.
Affiliation
  • Wallace AN; From the Mallinckrodt Institute of Radiology (A.N.W., T.J.H., M.V.F., J.W.J.) wallacea@mir.wustl.edu.
  • Hillen TJ; From the Mallinckrodt Institute of Radiology (A.N.W., T.J.H., M.V.F., J.W.J.).
  • Friedman MV; From the Mallinckrodt Institute of Radiology (A.N.W., T.J.H., M.V.F., J.W.J.).
  • Zohny ZS; Department of Neurosurgery (Z.S.Z., B.H.S.).
  • Stephens BH; Department of Neurosurgery (Z.S.Z., B.H.S.).
  • Greco SC; Division of Comparative Medicine (S.C.G., M.R.T.), Washington University, St Louis, Missouri.
  • Talcott MR; Division of Comparative Medicine (S.C.G., M.R.T.), Washington University, St Louis, Missouri.
  • Jennings JW; From the Mallinckrodt Institute of Radiology (A.N.W., T.J.H., M.V.F., J.W.J.).
AJNR Am J Neuroradiol ; 38(8): 1653-1659, 2017 Aug.
Article in En | MEDLINE | ID: mdl-28572150
ABSTRACT
BACKGROUND AND

PURPOSE:

Despite the growing use of percutaneous ablation therapy for the treatment of metastatic spine disease, several issues have yet to be fully addressed. Our aims were to determine whether the vertebral body cortex protects against ablation-induced spinal cord injury; correlate radiofrequency, cryo-, and microwave ablation parameters with resulting spinal ablation zone dimensions and describe normal spinal marrow postablation changes on MR imaging. MATERIALS AND

METHODS:

Ten thoracolumbar vertebrae in 3 sheep were treated with radiofrequency ablation, cryoablation, or microwave ablation under fluoroscopic guidance. Technique parameters were chosen to produce ablation zones that exceeded the volume of the vertebral bodies in sheep 1 and were confined to the vertebrae in sheep 2 and 3. Expected ablation zone dimensions were based on data provided by the device manufacturers. Postablation MR imaging was performed at 48 hours (sheep 1) or 7 days (sheep 2 and 3).

RESULTS:

In sheep 1, cryoablation and microwave ablations extended into the spinal canal and caused histologically confirmed neurologic injury, but radiofrequency ablation did not. The mean difference between the lengths of the radiofrequency ablation zone dimensions measured on gross pathology compared with those expected was 9.6 ± 4.1 mm. The gross pathologic cryo- and microwave ablation zone dimensions were within 1 mm of those expected. All modalities produced a nonenhancing ablation zone with a rim of enhancement, corresponding histologically to marrow necrosis and hemorrhagic congestion.

CONCLUSIONS:

An intact cortex appears to protect against radiofrequency ablation-induced spinal cord injury, but not against non-impedance-based modalities. Ablation dimensions produced by microwave and cryoablation are similar to those expected, while radiofrequency ablation dimensions are smaller. Ablation of normal marrow produces a rim of enhancement at the margin of the ablation zone on MR imaging.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Spine / Neurosurgical Procedures / Ablation Techniques Type of study: Diagnostic_studies / Etiology_studies / Guideline / Prognostic_studies Limits: Animals Language: En Journal: AJNR Am J Neuroradiol Year: 2017 Document type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Spine / Neurosurgical Procedures / Ablation Techniques Type of study: Diagnostic_studies / Etiology_studies / Guideline / Prognostic_studies Limits: Animals Language: En Journal: AJNR Am J Neuroradiol Year: 2017 Document type: Article