Your browser doesn't support javascript.
loading
Passive Picoinjection Enables Controlled Crystallization in a Droplet Microfluidic Device.
Li, Shunbo; Zeng, Muling; Gaule, Thembaninkosi; McPherson, Michael J; Meldrum, Fiona C.
Affiliation
  • Li S; School of Chemistry, University of Leeds, Leeds, LS2 9JT, UK.
  • Zeng M; School of Chemistry, University of Leeds, Leeds, LS2 9JT, UK.
  • Gaule T; Astbury Centre for Structural Molecular Biology, School of Molecular and Cellular Biology, University of Leeds, Leeds, LS2 9JT, UK.
  • McPherson MJ; Astbury Centre for Structural Molecular Biology, School of Molecular and Cellular Biology, University of Leeds, Leeds, LS2 9JT, UK.
  • Meldrum FC; School of Chemistry, University of Leeds, Leeds, LS2 9JT, UK.
Small ; 13(41)2017 11.
Article in En | MEDLINE | ID: mdl-28873281
Segmented flow microfluidic devices offer an attractive means of studying crystallization processes. However, while they are widely employed for protein crystallization, there are few examples of their use for sparingly soluble compounds due to problems with rapid device fouling and irreproducibility over longer run-times. This article presents a microfluidic device which overcomes these issues, as this is constructed around a novel design of "picoinjector" that facilitates direct injection into flowing droplets. Exploiting a Venturi junction to reduce the pressure within the droplet, it is shown that passive injection of solution from a side-capillary can be achieved in the absence of an applied electric field. The operation of this device is demonstrated for calcium carbonate, where highly reproducible results are obtained over long run-times at high supersaturations. This compares with conventional devices that use a Y-junction to achieve solution loading, where in-channel precipitation of calcium carbonate occurs even at low supersaturations. This work not only opens the door to the use of microfluidics to study the crystallization of low solubility compounds, but the simple design of a passive picoinjector will find wide utility in areas including multistep reactions and investigation of reaction dynamics.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Small Journal subject: ENGENHARIA BIOMEDICA Year: 2017 Document type: Article Country of publication: Alemania

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Small Journal subject: ENGENHARIA BIOMEDICA Year: 2017 Document type: Article Country of publication: Alemania