Your browser doesn't support javascript.
loading
The New Generation Immunochemical Test for Fecal Occult Blood (ColonView Quick Test) Shows a High Diagnostic Accuracy in Colorectal Cancer Detection.
Meklin, Jannica; Eskelinen, Maaret; Guimaraes, Denise Peixoto; Selander, Tuomas; Inkinen, Jenni; Tiusanen, Tapani; Syrjänen, Kari; Eskelinen, Matti.
Affiliation
  • Meklin J; Department of Surgery, Kuopio University Hospital and School of Medicine, University of Eastern Finland, Kuopio, Finland.
  • Eskelinen M; Department of Surgery, Kuopio University Hospital and School of Medicine, University of Eastern Finland, Kuopio, Finland.
  • Guimaraes DP; Molecular Oncology Research Center, Barretos Cancer Hospital, Barretos, Brazil.
  • Selander T; Science Service Center, Kuopio University Hospital and School of Medicine, University of Eastern Finland, Kuopio, Finland.
  • Inkinen J; Department of Clinical Research, Biohit Oyj, Helsinki, Finland.
  • Tiusanen T; Department of Clinical Research, Biohit Oyj, Helsinki, Finland.
  • Syrjänen K; Molecular Oncology Research Center, Barretos Cancer Hospital, Barretos, Brazil.
  • Eskelinen M; SMW Consultants, Ltd., Kaarina, Finland.
Anticancer Res ; 41(10): 5071-5079, 2021 Oct.
Article in En | MEDLINE | ID: mdl-34593457
ABSTRACT
BACKGROUND/

AIM:

The present study compared the accuracy of visually analyzed (VA) and automatically analyzed (AA) ColonView (CV) quick test; a new-generation immunochemical test (FIT) for Hb and Hb/Hp (Biohit Oyj, Helsinki, Finland) in subjects participating in colorectal cancer (CRC) detection in Brazil. A traditional gFOBT test (HemoccultSENSA) was used as a reference. PATIENTS AND

METHODS:

A cohort of 368 colonoscopy-referral patients were asked to collect 3 consecutive fecal samples, to be analysed by both assays (CV, SENSA).

RESULTS:

In receiver operating characteristic (ROC) analysis for the AA reading, the optimal cut-off value for CV Hb AA (test AA 3) was ≥117 and that for CV Hb/Hp AA (test AA 4) was ≥48. In the hierarchical summary receiver operating characteristic (HSROC) analysis for pooled accuracy of CV with AA and VA reading, the AUC values for i) VA and ii) AA were as follows i) AUC=0.859 (95%CI=0.839-0.879), ii) AUC=0.931 (95%CI=0.920-0.942). The difference between these AUC values (Roccomp analysis) was statistically significant (p=0.0024).

CONCLUSION:

The present study confirms the previous studies on the applicability of the ColonView quick test (a new-generation FIT) in CRC screening.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Immunochemistry / Colorectal Neoplasms / Colonoscopy / Early Detection of Cancer / Occult Blood Type of study: Diagnostic_studies / Prognostic_studies / Screening_studies Limits: Humans Language: En Journal: Anticancer Res Year: 2021 Document type: Article Affiliation country: Finlandia

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Immunochemistry / Colorectal Neoplasms / Colonoscopy / Early Detection of Cancer / Occult Blood Type of study: Diagnostic_studies / Prognostic_studies / Screening_studies Limits: Humans Language: En Journal: Anticancer Res Year: 2021 Document type: Article Affiliation country: Finlandia