Your browser doesn't support javascript.
loading
T1D -weighted ihMT imaging - Part II. Investigating the long- and short-T1D components correlation with myelin content. Comparison with R1 and the macromolecular proton fraction.
Hertanu, Andreea; Soustelle, Lucas; Buron, Julie; Le Priellec, Julie; Cayre, Myriam; Le Troter, Arnaud; Varma, Gopal; Alsop, David C; Durbec, Pascale; Girard, Olivier M; Duhamel, Guillaume.
Affiliation
  • Hertanu A; Aix Marseille Univ, CNRS, CRMBM, Marseille, France.
  • Soustelle L; APHM, Hôpital Universitaire Timone, CEMEREM, Marseille, France.
  • Buron J; Aix Marseille Univ, CNRS, CRMBM, Marseille, France.
  • Le Priellec J; APHM, Hôpital Universitaire Timone, CEMEREM, Marseille, France.
  • Cayre M; Aix Marseille Univ, CNRS, CRMBM, Marseille, France.
  • Le Troter A; APHM, Hôpital Universitaire Timone, CEMEREM, Marseille, France.
  • Varma G; Aix Marseille Univ, CNRS, IBDM, Marseille, France.
  • Alsop DC; Aix Marseille Univ, CNRS, IBDM, Marseille, France.
  • Durbec P; Aix Marseille Univ, CNRS, IBDM, Marseille, France.
  • Girard OM; Aix Marseille Univ, CNRS, CRMBM, Marseille, France.
  • Duhamel G; APHM, Hôpital Universitaire Timone, CEMEREM, Marseille, France.
Magn Reson Med ; 87(5): 2329-2346, 2022 05.
Article in En | MEDLINE | ID: mdl-35001427
ABSTRACT

PURPOSE:

To investigate the long- and short-T1D components correlation with myelin content using inhomogeneous magnetization transfer (ihMT) high-pass and band-pass T1D -filters and to compare ihMT, R1 , and the macromolecular proton fraction (MPF) for myelin specific imaging.

METHODS:

The 3D ihMT rapid gradient echo (ihMTRAGE) sequences with increasing switching times (Δt) were used to derive ihMT high-pass T1D -filters with increasing T1D cutoff values and an ihMT band-pass T1D -filter for components in the 100 µs to 1 ms range. 3D spoiled gradient echo quantitative MT (SPGR-qMT) protocols were used to derive R1 and MPF maps. The specificity of R1 , MPF, and ihMT T1D -filters was evaluated by comparison with two histological reference techniques for myelin imaging.

RESULTS:

The higher contribution of long-T1D s as compared to the short components as Δt got longer led to an increase in the specificity to myelination. In contrast, focusing on the signal originating from a narrow range of short-T1D s (< 1 ms) as isolated by the band-pass T1D -filter led to lower specificity. In addition, the significantly lower r2 correlation coefficient of the band-pass T1D -filter suggests that the origin of short-T1D components is mostly associated with non-myelin protons. Also, the important contribution of short-T1D s to the estimated MPF, explains its low specificity to myelination as compared to the ihMT high-pass T1D -filters.

CONCLUSION:

Long-T1D components imaging by means of ihMT high-pass T1D -filters is proposed as an MRI biomarker for myelin content. Future studies should enable the investigation of the sensitivity of ihMT T1D -filters for demyelinating processes.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: White Matter / Myelin Sheath Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2022 Document type: Article Affiliation country: Francia

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: White Matter / Myelin Sheath Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2022 Document type: Article Affiliation country: Francia