Real-time tissue perfusion assessment using fluorescence imaging topography scanning system: A preclinical investigation.
Quang, Tri T; Wei, Chen F; Walsh, Susan A; Papay, Francis A; Liu, Yang.
Affiliation
  • Quang TT; Department of Electrical and Computer Engineering, The University of Iowa, Iowa City, Iowa, USA.
  • Wei CF; University of Iowa Technology Institute, The University of Iowa, Iowa City, Iowa, USA.
  • Walsh SA; Center for Bioinformatics and Computational Biology, The University of Iowa, Iowa City, Iowa, USA.
  • Papay FA; Iowa Informatics Initiative, The University of Iowa, Iowa City, Iowa, USA.
  • Liu Y; Iowa Institute for Biomedical Imaging, The University of Iowa, Iowa City, Iowa, USA.
Lasers Surg Med ; 54(7): 994-1001, 2022 09.
Article in En | MEDLINE | ID: mdl-35652907
ABSTRACT
BACKGROUND AND

OBJECTIVES:

We previously developed a real-time fluorescence imaging topography scanning (RFITS) system for intraoperative multimodal imaging, image-guided surgery, and dynamic surgical navigation. The RFITS can capture intraoperative fluorescence, color reflectance, and surface topography concurrently and offers accurate registration of multimodal images. The RFITS prototype is a promising system for multimodal image guidance and intuitive 3D visualization. In the current study, we investigated the capability of the RFITS system in intraoperative fluorescence vascular angiography for real-time assessment of tissue perfusion. STUDY DESIGN/MATERIALS AND

METHODS:

We conducted ex vivo imaging of fluorescence perfusion in a soft casting life-sized human brain phantom. Indocyanine green (ICG) solutions diluted in dimethyl sulfoxide (DMSO) and human serum were injected into the brain phantom through the vessel simulating tube (2 ± 0.2 mm inner diameter) by an adjustable flow peristaltic pump. To demonstrate the translational potential of the system, an ICG/DMSO solution was perfused into blood vessels of freshly harvested porcine ears (n = 9, inner diameter from 0.56 to 1.27 mm). We subsequently performed in vivo imaging of fluorescence-perfused vascular structures in rodent models (n = 10). 5 mg/ml ICG solutions prepared in sterile water were injected via the lateral tail vein. All targets were imaged by the RFITS prototype at a working distance of 350-400 mm.

RESULTS:

3D visualization of 10 µg/ml ICG-labeled continuous moving serum in the brain phantom was obtained at an average signal-to-background ratio (SBR) of 1.74 ± 0.03. The system was able to detect intravenously diffused fluorescence in porcine tissues with an average SBR of 2.23 ± 0.22. The RFITS prototype provided real-time monitoring of tissue perfusion in rats after intravenous (IV) administration of ICG. The maximum fluorescence intensity (average SBR = 1.94 ± 0.16, p < 0.001) was observed at Tpeak of ~30 seconds after the ICG signal was first detected (average SBR = 1.19 ± 0.13, p < 0.01).

CONCLUSIONS:

We have conducted preclinical studies to demonstrate the feasibility of applying the RFITS system in real-time fluorescence angiography and tissue perfusion assessment. Our system provides fluorescence/color composite images for intuitive visualization of tissue perfusion with 3D perception. The findings pave the way for future clinical translation.