Phantom study of stereotactic radioablation for ventricular tachycardia (STRA-MI-VT) using Cyberknife Synchrony Respiratory Tracking System with a single fiducial marker.
Piccolo, C; Vigorito, S; Rondi, E; Piperno, G; Ferrari, A; Pepa, M; Riva, G; Durante, S; Conte, E; Catto, V; Andreini, D; Carbucicchio, C; Jereczek-Fossa, B A; Pompilio, G; Orecchia, R; Cattani, F.
Affiliation
  • Piccolo C; Unit of Medical Physics, IEO European Institute of Oncology, IRCCS, Milan, Italy. Electronic address: consiglia.piccolo@chuv.ch.
  • Vigorito S; Unit of Medical Physics, IEO European Institute of Oncology, IRCCS, Milan, Italy.
  • Rondi E; Unit of Medical Physics, IEO European Institute of Oncology, IRCCS, Milan, Italy.
  • Piperno G; Division of Radiation Oncology, IEO European Institute of Oncology, IRCCS, Milan, Italy.
  • Ferrari A; Division of Radiation Oncology, IEO European Institute of Oncology, IRCCS, Milan, Italy.
  • Pepa M; Division of Radiation Oncology, IEO European Institute of Oncology, IRCCS, Milan, Italy.
  • Riva G; Division of Radiation Oncology, IEO European Institute of Oncology, IRCCS, Milan, Italy.
  • Durante S; Division of Radiation Oncology, IEO European Institute of Oncology, IRCCS, Milan, Italy.
  • Conte E; Cardiovascular Imaging Department, Centro Cardiologico Monzino IRCCS, Milan, Italy; Department of Biomedical Sciences for Health, University of Milan, Milan, Italy.
  • Catto V; Department of Clinical Electrophysiology and Cardiac Pacing, Centro Cardiologico Monzino IRCCS, Milan, Italy; Department of Electronics, Information and Biomedical Engineering, Politecnico di Milano, Milan, Italy.
  • Andreini D; Cardiovascular Imaging Department, Centro Cardiologico Monzino IRCCS, Milan, Italy; Department of Biomedical and Clinical Sciences "Luigi Sacco", University of Milan, Milan, Italy.
  • Carbucicchio C; Department of Clinical Electrophysiology and Cardiac Pacing, Centro Cardiologico Monzino IRCCS, Milan, Italy.
  • Jereczek-Fossa BA; Division of Radiation Oncology, IEO European Institute of Oncology, IRCCS, Milan, Italy; Department of Oncology and Hemato-Oncology, University of Milan, Milan, Italy.
  • Pompilio G; Scientific Directorate, Centro Cardiologico Monzino IRCCS, Milan, Italy; Department of Biomedical, Surgical and Dental Sciences, University of Milan, Milan, Italy.
  • Orecchia R; Scientific Directorate, IEO European Institute of Oncology IRCCS, Milan, Italy.
  • Cattani F; Unit of Medical Physics, IEO European Institute of Oncology, IRCCS, Milan, Italy.
Phys Med ; 100: 135-141, 2022 Aug.
Article in En | MEDLINE | ID: mdl-35816942
ABSTRACT

PURPOSE:

Within the STRA-MI-VT phase Ib/II trial (NCT04066517), the aim of this phantom study was to explore the feasibility of Cyberknife treatments on cardiac lesions by tracking as a single marker the lead tip of an implantable cardioverter defibrillator. The residual displacement of the lesion during the tracking was studied, planning margins were found and the dosimetric accuracy of the treatment was checked. MATERIALS AND

METHODS:

A lead was inserted into a phantom (EasyCube phantom, Sun Nuclear Co, USA) and then placed on the translating ExacTrac Gating System (BrainLAB AG, Germany). The phantom was rotated, a virtual lesion was identified and its displacement during the tracking was studied. Two plans were compared, calculated on the unrotated volume and on the envelope of the unrotated and the rotated volumes. The plans were delivered using the Cyberknife System (Accuray Inc, USA) and their dosimetric accuracy verified by gamma analysis with gafchromic films.

RESULTS:

The residual margin increases enhancing the distance between the lead and the lesion. It is 4 mm for distance 0 cm and 5 mm for distance 5 cm. The coverage is reduced by 3.8% (interquartile range 2.5%-4.7%) when the dose is prescribed on the unrotated volume. All treatment plans are accurate and 3% 3 mm gamma analysis results are greater than 94%.

CONCLUSIONS:

Results showed that tracking with a single marker is feasible considering adequate residual planning margins. The volumes could be further reduced by using additional markers, for example by placing them on the patient's skin.