Evaluation of MetaSystems Automated Fluorescent Microscopy System for the Machine-Assisted Detection of Acid-Fast Bacilli in Clinical Samples.
Tomasello, Gianna; Foroughi, Farnaz; Padron, Danielle; Moreno, Angel; Banaei, Niaz.
Affiliation
  • Tomasello G; Clinical Microbiology Laboratory, Stanford Health Care, Palo Alto, California, USA.
  • Foroughi F; Clinical Microbiology Laboratory, Stanford Health Care, Palo Alto, California, USA.
  • Padron D; Clinical Microbiology Laboratory, Stanford Health Care, Palo Alto, California, USA.
  • Moreno A; Department of Pathology, Stanford University School of Medicinegrid.471392.a, Stanford, California, USA.
  • Banaei N; Clinical Microbiology Laboratory, Stanford Health Care, Palo Alto, California, USA.
J Clin Microbiol ; 60(10): e0113122, 2022 Oct 19.
Article in En | MEDLINE | ID: mdl-36121216
ABSTRACT
Manual reading of fluorescent acid-fast bacilli (AFB) microscopy slides is time-intensive and technically demanding. The aim of this study was to evaluate the accuracy of MetaSystems' automated fluorescent AFB slide scanner and analyzer. Auramine O-stained slides corresponding to 133 culture-positive and 363 culture-negative respiratory (n = 284), tissue (n = 120), body fluid (n = 81), and other (n = 11) sources were evaluated with the MetaSystems Mycobacteria Scanner running the NEON Metafer AFB Module. The sensitivity and specificity of the MetaSystems platform was measured as a standalone diagnostic and as an assistant to technologists to review positive images. Culture results were used as the reference method. The MetaSystems platform failed to scan 57 (11.5%) slides. The MetaSystems platform used as a standalone had a sensitivity of 97.0% (129/133; 95% CI 92.5 to 99.2) and specificity of 12.7% (46/363; 95% CI 9.4 to 16.5). When positive scans were used to assist technologists, the MetaSystems platform had a sensitivity of 70.7% (94/133; 95% CI 62.2 to 78.3) and specificity of 89.0% (323/363; 95% CI 85.3 to 92.0). The manual microscopy method had a sensitivity of 79.7% (106/133; 95% CI 71.9 to 86.2) and specificity of 98.6% (358/363; 95% CI 96.8 to 99.6). The sensitivity of the MetaSystems platform was not impacted by smear grade or mycobacterial species. The majority (70.3%) of false positive smears had ≥2+ smear results with the MetaSystems platform. Further performance improvements are needed before the MetaSystems' automated fluorescent AFB slide reader can be used to assist microscopist in the clinical laboratory.