Long-term safety and efficacy of left atrial appendage closure in patients with small appendage orifices measured with transesophageal echocardiography.
Wang, Binhao; Wang, Zhao; Chu, Huimin; He, Bin; Fu, Guohua; Feng, Mingjun; Du, Xianfeng; Liu, Jing; Yu, Yibo.
Affiliation
  • Wang B; Arrhythmia Center, Ningbo First Hospital, Ningbo, China.
  • Wang Z; Key Laboratory of Precision Medicine for Atherosclerotic Diseases of Zhejiang Province, Ningbo, China.
  • Chu H; Department of Ultrasonography, Ningbo First Hospital, Ningbo, China.
  • He B; Arrhythmia Center, Ningbo First Hospital, Ningbo, China.
  • Fu G; Key Laboratory of Precision Medicine for Atherosclerotic Diseases of Zhejiang Province, Ningbo, China.
  • Feng M; Arrhythmia Center, Ningbo First Hospital, Ningbo, China.
  • Du X; Key Laboratory of Precision Medicine for Atherosclerotic Diseases of Zhejiang Province, Ningbo, China.
  • Liu J; Arrhythmia Center, Ningbo First Hospital, Ningbo, China.
  • Yu Y; Key Laboratory of Precision Medicine for Atherosclerotic Diseases of Zhejiang Province, Ningbo, China.
Clin Cardiol ; 46(2): 134-141, 2023 Feb.
Article in En | MEDLINE | ID: mdl-36378742
ABSTRACT

BACKGROUND:

The Watchman device is the most widely used occluder but is indicated in atrial fibrillation (AF) patients with a maximal left atrial appendage (LAA) orifice diameter between 17 and 31 mm. We aimed to compare the long-term safety and efficacy of left atrial appendage closure (LAAC) between patients with a small LAA (<17 mm) and those with an indicated LAA (17-31 mm) measured by transesophageal echocardiography (TEE).

METHODS:

A total of 369 AF patients treated with LAAC between March 2015 and February 2019 were included and divided into two groups based on the maximal LAA orifice diameter measured by TEE small LAA group (n = 22) and indicated LAA group (n = 347). Periprocedural complications and long-term clinical outcomes were compared.

RESULTS:

The Watchman device was successfully implanted in all patients. Mean device compression was higher in the small LAA group. Four patients (1.2%) in the indicated LAA group experienced pericardial effusion, and none experienced pericardial effusion in the small LAA group. Device-related thrombus was detected in one (4.5%) patient in the small LAA group and five (1.4%) in the indicated LAA group (p = .310). After a mean follow-up period of 4.1 ± 1.6 years, one patient in the small LAA group (4.5%; 1.1/100 person-years) and four in the indicated LAA group (1.2%; 0.3/100 person-years) suffered an ischemic stroke (p = .266).

CONCLUSIONS:

The safety and efficacy of LAAC with the Watchman device were comparable between patients with small and indicated LAA orifice diameters measured by TEE.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pericardial Effusion / Atrial Fibrillation / Atrial Appendage Limits: Humans Language: En Journal: Clin Cardiol Year: 2023 Document type: Article Affiliation country: China

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pericardial Effusion / Atrial Fibrillation / Atrial Appendage Limits: Humans Language: En Journal: Clin Cardiol Year: 2023 Document type: Article Affiliation country: China