Your browser doesn't support javascript.
loading
In-vivo quality assurance of dynamic tumor tracking (DTT) for liver SABR using EPID images.
Rostamzadeh, Maryam; Luchka, Kurt; Ma, Roy; Liu, Mitchell; Dunne, Emma; Camborde, Marie-Laure; Karan, Tania; Mestrovic, Ante; Bergman, Alanah.
Affiliation
  • Rostamzadeh M; Department of Physics and Astronomy, University of British Columbia, Vancouver, BC, Canada.
  • Luchka K; Medical Physics Department, BC Cancer-Vancouver, Vancouver, Canada.
  • Ma R; Radiation Oncology Department, BC Cancer-Vancouver, Vancouver, Canada.
  • Liu M; Radiation Oncology Department, BC Cancer-Vancouver, Vancouver, Canada.
  • Dunne E; Radiation Oncology Department, BC Cancer-Vancouver, Vancouver, Canada.
  • Camborde ML; Medical Physics Department, BC Cancer-Vancouver, Vancouver, Canada.
  • Karan T; Medical Physics Department, BC Cancer-Vancouver, Vancouver, Canada.
  • Mestrovic A; Medical Physics Department, BC Cancer-Vancouver, Vancouver, Canada.
  • Bergman A; Medical Physics Department, BC Cancer-Vancouver, Vancouver, Canada.
J Appl Clin Med Phys ; 24(7): e13969, 2023 Jul.
Article in En | MEDLINE | ID: mdl-36995913
ABSTRACT

PURPOSE:

To assess dynamic tumor tracking (DTT) target localization uncertainty for in-vivo marker-based stereotactic ablative radiotherapy (SABR) treatments of the liver using electronic-portal-imaging-device (EPID) images. The Planning Target Volume (PTV) margin contribution for DTT is estimated.

METHODS:

Phantom and patient EPID images were acquired during non-coplanar 3DCRT-DTT delivered on a Vero4DRT linac. A chain-code algorithm was applied to detect Multileaf Collimator (MLC)-defined radiation field edges. Gold-seed markers were detected using a connected neighbor algorithm. For each EPID image, the absolute differences between the measured center-of-mass (COM) of the markers relative to the aperture-center (Tracking Error, (ET )) was reported in pan, tilt, and 2D-vector directions at the isocenter-plane. PHANTOM STUDY An acrylic cube phantom implanted with gold-seed markers was irradiated with non-coplanar 3DCRT-DTT beams and EPID images collected. Patient Study Eight liver SABR patients were treated with non-coplanar 3DCRT-DTT beams. All patients had three to four implanted gold-markers. In-vivo EPID images were analyzed.

RESULTS:

Phantom Study On the 125 EPID images collected, 100% of the markers were identified. The average ± SD of ET were 0.24 ± 0.21, 0.47 ± 0.38, and 0.58 ± 0.37 mm in pan, tilt and 2D directions, respectively. Patient Study Of the 1430 EPID patient images acquired, 78% had detectable markers. Over all patients, the average ± SD of ET was 0.33 ± 0.41 mm in pan, 0.63 ± 0.75 mm in tilt and 0.77 ± 0.80 mm in 2D directions The random 2D-error, σ, for all patients was 0.79 mm and the systematic 2D-error, Σ, was 0.20 mm. Using the Van Herk margin formula 1.1 mm planning target margin can represent the marker based DTT uncertainty.

CONCLUSIONS:

Marker-based DTT uncertainty can be evaluated in-vivo on a field-by-field basis using EPID images. This information can contribute to PTV margin calculations for DTT.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Radiosurgery / Radiotherapy, Conformal / Neoplasms Limits: Humans Language: En Journal: J Appl Clin Med Phys Journal subject: BIOFISICA Year: 2023 Document type: Article Affiliation country: Canadá

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Radiosurgery / Radiotherapy, Conformal / Neoplasms Limits: Humans Language: En Journal: J Appl Clin Med Phys Journal subject: BIOFISICA Year: 2023 Document type: Article Affiliation country: Canadá