[Evaluation of Low-contrast Detectability Using the Digital Phantom Creation Tool in the Late Arterial Phase to Detect Liver Mass Lesions].
Harada, Kohei; Imai, Tatsuya; Ohashi, Yoshiya; Chiba, Ayaka; Numasawa, Kanako; Hayasaka, Shun; Omori, Go.
Affiliation
  • Harada K; Division of Radiology, Sapporo Medical University Hospital.
  • Imai T; Division of Radiology, Sapporo Medical University Hospital.
  • Ohashi Y; Division of Radiology, Sapporo Medical University Hospital.
  • Chiba A; Division of Radiology, Sapporo Medical University Hospital.
  • Numasawa K; Division of Radiology, Sapporo Medical University Hospital.
  • Hayasaka S; Division of Radiology, Sapporo Medical University Hospital.
  • Omori G; Division of Radiology, Sapporo Medical University Hospital.
Nihon Hoshasen Gijutsu Gakkai Zasshi ; 79(7): 674-681, 2023 Jul 20.
Article in Ja | MEDLINE | ID: mdl-37286500
ABSTRACT

PURPOSE:

Late arterial phase images of SD 8, SD 10, and SD 12 were acquired in the 3-phase dynamic study of the liver in combination with hybrid iterative reconstruction. We evaluated the low-contrast detectability by adding a simulated tumor to these images and aimed to formulate a standard image quality.

METHODS:

We prepared images with and without signal for 60 series of 20 samples, each with 3 image quality types (total 120 series). The continuous confidence method by 10 observers detected 60 simulated tumors.

RESULTS:

The detection sensitivities were 0.765, 0.785, and 0.260 for SD 8, SD 10, and SD 12, respectively (p<0.001) with no significantly different specificities, and the areas under the curve were 0.901, 0.892, and 0.616 (p<0.001), respectively. The simulated mass detection rates were 74.5%, 75.0%, and 21.5% for SD 8, SD 10, and SD 12, respectively (p<0.001), and the intraclass correlation coefficients, which indicate interobserver reliability, were 0.697 at SD 10 without signal, and SD 12 without a signal significantly dropped to 0.185.

CONCLUSION:

Therefore, SD 12 images increase the possibility of overlooking lesions. Hence, image quality in the late arterial phase should be SD 10 or less.