Optical detection of infiltration during peripheral intravenous infusion in neonates.
D'Andrea, Vito; Prontera, Giorgia; Carlino, Riccardo; Di Trani, Helena; Carlettini, Ilaria; Pittiruti, Mauro; Vento, Giovanni.
Affiliation
  • D'Andrea V; Neonatology Unit, Department of Woman and Child Health and Public Health, Fondazione Policlinico Universitario Agostino Gemelli IRCCS, Rome, Italy.
  • Prontera G; Neonatology Unit, Department of Woman and Child Health and Public Health, Fondazione Policlinico Universitario Agostino Gemelli IRCCS, Rome, Italy.
  • Carlino R; Neonatology Unit, Department of Woman and Child Health and Public Health, Fondazione Policlinico Universitario Agostino Gemelli IRCCS, Rome, Italy.
  • Di Trani H; Neonatology Unit, Department of Woman and Child Health and Public Health, Fondazione Policlinico Universitario Agostino Gemelli IRCCS, Rome, Italy.
  • Carlettini I; Neonatology Unit, Department of Woman and Child Health and Public Health, Fondazione Policlinico Universitario Agostino Gemelli IRCCS, Rome, Italy.
  • Pittiruti M; Department of Surgery, Fondazione Policlinico Gemelli, Rome, Italy.
  • Vento G; Neonatology Unit, Department of Woman and Child Health and Public Health, Fondazione Policlinico Universitario Agostino Gemelli IRCCS, Rome, Italy.
J Vasc Access ; : 11297298231177723, 2023 Jul 13.
Article in En | MEDLINE | ID: mdl-37439226
ABSTRACT

BACKGROUND:

Infiltration and extravasation are common complications during peripheral intravenous infusion in the neonatal intensive care unit, and diagnosis is usually clinical, by inspection and palpation. Delay in diagnosis due to poor surveillance or misinterpretation of clinical signs may carry serious damage to the tissues of the neonate. Recently, a novel technology based on optical detection of infiltration has become available.

METHODS:

We have studied two groups of term and preterm infants receiving non-vesicant intravenous infusions by the peripheral route (24G short peripheral cannulas), and we evaluated the incidence of infiltration. In the first group, we have compared the clinical detection of infiltration versus the detection obtained by a novel optical device, blinding the alarms of the device. In the second group, the comparison was carried out without blinding the alarms.

RESULTS:

Of the neonates included in this study, 60% were female, 86% had a gestational age <37 weeks (34 ± 2.5 weeks) and a mean birth weight of 2.08 ± 0.4 g. Total recorded monitoring time was 1318 h and average monitoring time for each short peripheral cannula was 26.4 h. The incidence of infiltration was 80%, most of them having a Millam score <2. The novel device showed an overall sensitivity of 88.9% in detecting infiltration. Specificity was 84.4%, as evaluated assuming as standard the clinical diagnosis. Interestingly, in cases of documented infiltration, the event was diagnosed by the optical device approximately 6 h before the clinical diagnosis.

CONCLUSIONS:

Continuous monitoring of the insertion site, as automatically ensured by this novel optical device, may play a complementary role in early detection of infiltration, even if the percentage of false positives and false negatives suggests that periodic clinical assessment by expert nurses cannot be omitted.