Your browser doesn't support javascript.
loading
Effects of skin tone on photoacoustic imaging and oximetry.
Else, Thomas R; Hacker, Lina; Gröhl, Janek; Bunce, Ellie V; Tao, Ran; Bohndiek, Sarah E.
Affiliation
  • Else TR; University of Cambridge, CRUK Cambridge Institute, Cambridge, United Kingdom.
  • Hacker L; University of Cambridge, Department of Physics, Cambridge, United Kingdom.
  • Gröhl J; University of Cambridge, CRUK Cambridge Institute, Cambridge, United Kingdom.
  • Bunce EV; University of Cambridge, Department of Physics, Cambridge, United Kingdom.
  • Tao R; University of Cambridge, CRUK Cambridge Institute, Cambridge, United Kingdom.
  • Bohndiek SE; University of Cambridge, Department of Physics, Cambridge, United Kingdom.
J Biomed Opt ; 29(Suppl 1): S11506, 2024 Jan.
Article in En | MEDLINE | ID: mdl-38125716
ABSTRACT

Significance:

Photoacoustic imaging (PAI) provides contrast based on the concentration of optical absorbers in tissue, enabling the assessment of functional physiological parameters such as blood oxygen saturation (sO2). Recent evidence suggests that variation in melanin levels in the epidermis leads to measurement biases in optical technologies, which could potentially limit the application of these biomarkers in diverse populations.

Aim:

To examine the effects of skin melanin pigmentation on PAI and oximetry.

Approach:

We evaluated the effects of skin tone in PAI using a computational skin model, two-layer melanin-containing tissue-mimicking phantoms, and mice of a consistent genetic background with varying pigmentations. The computational skin model was validated by simulating the diffuse reflectance spectrum using the adding-doubling method, allowing us to assign our simulation parameters to approximate Fitzpatrick skin types. Monte Carlo simulations and acoustic simulations were run to obtain idealized photoacoustic images of our skin model. Photoacoustic images of the phantoms and mice were acquired using a commercial instrument. Reconstructed images were processed with linear spectral unmixing to estimate blood oxygenation. Linear unmixing results were compared with a learned unmixing approach based on gradient-boosted regression.

Results:

Our computational skin model was consistent with representative literature for in vivo skin reflectance measurements. We observed consistent spectral coloring effects across all model systems, with an overestimation of sO2 and more image artifacts observed with increasing melanin concentration. The learned unmixing approach reduced the measurement bias, but predictions made at lower blood sO2 still suffered from a skin tone-dependent effect.

Conclusion:

PAI demonstrates measurement bias, including an overestimation of blood sO2, in higher Fitzpatrick skin types. Future research should aim to characterize this effect in humans to ensure equitable application of the technology.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Skin Pigmentation / Photoacoustic Techniques Limits: Animals / Humans Language: En Journal: J Biomed Opt / J. biomed. opt. (Online) / Journal of biomedical optics (Online) Journal subject: ENGENHARIA BIOMEDICA / OFTALMOLOGIA Year: 2024 Document type: Article Affiliation country: Reino Unido Country of publication: Estados Unidos

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Skin Pigmentation / Photoacoustic Techniques Limits: Animals / Humans Language: En Journal: J Biomed Opt / J. biomed. opt. (Online) / Journal of biomedical optics (Online) Journal subject: ENGENHARIA BIOMEDICA / OFTALMOLOGIA Year: 2024 Document type: Article Affiliation country: Reino Unido Country of publication: Estados Unidos