Your browser doesn't support javascript.
loading
A sensorised surgical glove to improve training and detection of obstetric anal sphincter injury: A preclinical study on a pig model.
Jaufuraully, Shireen R; Salvadores Fernandez, Carmen; Abbas, Nadine; Desjardins, Adrien; Tiwari, Manish K; David, Anna L; Siassakos, Dimitrios.
Affiliation
  • Jaufuraully SR; Wellcome/EPSRC Centre for Interventional and Surgical Sciences, University College London, London, UK.
  • Salvadores Fernandez C; Elizabeth Garrett Anderson Institute for Women's Health, University College London, London, UK.
  • Abbas N; Wellcome/EPSRC Centre for Interventional and Surgical Sciences, University College London, London, UK.
  • Desjardins A; Nanoengineered Systems Laboratory, Mechanical Engineering, University College London, London, UK.
  • Tiwari MK; Elizabeth Garrett Anderson Institute for Women's Health, University College London, London, UK.
  • David AL; Wellcome/EPSRC Centre for Interventional and Surgical Sciences, University College London, London, UK.
  • Siassakos D; Department of Medical Physics and Biomedical Engineering, University College London, London, UK.
BJOG ; 131(8): 1129-1135, 2024 Jul.
Article in En | MEDLINE | ID: mdl-38247347
ABSTRACT

OBJECTIVE:

To create a sensorised surgical glove that can accurately identify obstetric anal sphincter injury to facilitate timely repair, reduce complications and aid training.

DESIGN:

Proof-of-concept.

SETTING:

Laboratory. SAMPLE Pig models.

METHODS:

Flexible triboelectric pressure/force sensors were mounted onto the fingertips of a routine surgical glove. The sensors produce a current when rubbed on materials of different characteristics which can be analysed. A per rectum examination was performed on the intact sphincter of pig cadavers, analogous to routine examination for obstetric anal sphincter injuries postpartum. An anal sphincter defect was created by cutting through the vaginal mucosa and into the external anal sphincter using a scalpel. The sphincter was then re-examined. Data and signals were interpreted. MAIN OUTCOME

MEASURES:

Sensitivity and specificity of the glove in detecting anal sphincter injury.

RESULTS:

In all, 200 examinations were performed. The sensors detected anal sphincter injuries in a pig model with sensitivities between 98% and 100% and a specificity of 100%. The current produced when examining an intact sphincter and sphincter with a defect was significantly different (p < 0.001).

CONCLUSION:

In this preliminary study, the sensorised glove accurately detected anal sphincter injury in a pig model. Future plans include its clinical translation, starting with an in-human study on postpartum women, to determine whether it can accurately detect different types of obstetric anal sphincter injury in vivo.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Anal Canal / Gloves, Surgical Type of study: Diagnostic_studies / Prognostic_studies Limits: Animals / Female / Humans / Pregnancy Language: En Journal: BJOG Journal subject: GINECOLOGIA / OBSTETRICIA Year: 2024 Document type: Article Country of publication: Reino Unido

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Anal Canal / Gloves, Surgical Type of study: Diagnostic_studies / Prognostic_studies Limits: Animals / Female / Humans / Pregnancy Language: En Journal: BJOG Journal subject: GINECOLOGIA / OBSTETRICIA Year: 2024 Document type: Article Country of publication: Reino Unido