Continuous timely monitoring of core temperature with two wearable devices in pediatric patients undergoing chemotherapy for cancer - a comparison study.
Koenig, Christa; Ammann, Roland A; Schneider, Christine; Wyss, Johanna; Roessler, Jochen; Brack, Eva.
Affiliation
  • Koenig C; Pediatric Hematology/Oncology, Department of Pediatrics, Inselspital, Bern University Hospital, University of Bern, CH-3010, Bern, Switzerland. christa.koenig@insel.ch.
  • Ammann RA; , StatConsult Ammann, Burgdorf, Switzerland.
  • Schneider C; Faculty of Medicine, University of Bern, CH-3010, Bern, Switzerland.
  • Wyss J; Pediatric Hematology/Oncology, Department of Pediatrics, Inselspital, Bern University Hospital, University of Bern, CH-3010, Bern, Switzerland.
  • Roessler J; Department of Pediatric Oncology and Hematology, University Children's Hospital Basel (UKBB), University of Basel, Basel, Switzerland.
  • Brack E; Pediatric Hematology/Oncology, Department of Pediatrics, Inselspital, Bern University Hospital, University of Bern, CH-3010, Bern, Switzerland.
Support Care Cancer ; 32(3): 188, 2024 Feb 24.
Article in En | MEDLINE | ID: mdl-38400942
ABSTRACT

PURPOSE:

Pediatric patients with cancer often develop chemotherapy-induced fever in neutropenia (FN), requiring emergency broad-spectrum antibiotics. Continuous temperature monitoring can lead to earlier FN detection and therapy with improved outcomes. We aimed to compare the feasibility of continuous core temperature monitoring with timely data availability between two wearable devices (WDs) in pediatric oncology patients undergoing chemotherapy.

METHODS:

In this prospective observational two-center study, 20 patients (median age 8 years) undergoing chemotherapy simultaneously wore two WDs (CORE®, Everion®) for 14 days. The predefined goal was core temperature recorded in sufficient quality and available within ≤ 30 min during ≥ 18/24 h for ≥ 7/14 days in more than 15 patients.

RESULTS:

More patients reached the goal with CORE® (n = 13) versus Everion® (n = 3) (difference, 50% p < 0.001). After correcting for the transmission bottleneck caused by two WDs transmitting via one gateway, these numbers increased (n = 15 versus n = 14; difference, 5%; p = 0.69). CORE® measurements corresponded better to ear temperatures (n = 528; mean bias, - 0.07 °C; mean absolute difference, 0.35 °C) than Everion® measurements (n = 532; - 1.06 °C; 1.10 °C). Acceptance rates for the WDs were 95% for CORE® and 89% for Everion®.

CONCLUSION:

The CORE® fulfilled the predefined feasibility criterion (15 of 20 patients) after correction for transmission bottleneck, and the Everion® nearly fulfilled it. Continuous core temperature recording of good quality and with timely data availability was feasible from preschool to adolescent patients undergoing chemotherapy for cancer. These results encourage the design of randomized controlled trials on continuously monitored core temperature in pediatric patients. CLINICALTRIALS gov (NCT04914702) on June 7, 2021.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Wearable Electronic Devices / Neoplasms Limits: Adolescent / Child / Child, preschool / Humans Language: En Journal: Support Care Cancer Journal subject: NEOPLASIAS / SERVICOS DE SAUDE Year: 2024 Document type: Article Affiliation country: Suiza

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Wearable Electronic Devices / Neoplasms Limits: Adolescent / Child / Child, preschool / Humans Language: En Journal: Support Care Cancer Journal subject: NEOPLASIAS / SERVICOS DE SAUDE Year: 2024 Document type: Article Affiliation country: Suiza
...