Your browser doesn't support javascript.
loading
Evaluation of Stability and Accuracy Compared to the Westergren Method of ESR Samples Analyzed at VES-MATIC 5.
Lorubbio, Maria; Diamanti, Daniela; Ghiandai, Alessandro; Pieroni, Carolina; Bonini, Donatella; Pettinari, Massimiliano; Gorini, Gabriele; Bassi, Stefania; Meloni, Paola; Ognibene, Agostino.
Affiliation
  • Lorubbio M; Chemical-Clinical Analysis Laboratory, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Tuscany, Italy.
  • Diamanti D; DIESSE-Diagnostica Senese S.p.A., Strada dei Laghi, 35-39, 53035 Monteriggioni, Siena, Italy.
  • Ghiandai A; Chemical-Clinical Analysis Laboratory, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Tuscany, Italy.
  • Pieroni C; DIESSE-Diagnostica Senese S.p.A., Strada dei Laghi, 35-39, 53035 Monteriggioni, Siena, Italy.
  • Bonini D; Chemical-Clinical Analysis Laboratory, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Tuscany, Italy.
  • Pettinari M; Chemical-Clinical Analysis Laboratory, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Tuscany, Italy.
  • Gorini G; Chemical-Clinical Analysis Laboratory, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Tuscany, Italy.
  • Bassi S; Chemical-Clinical Analysis Laboratory, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Tuscany, Italy.
  • Meloni P; Chemical-Clinical Analysis Laboratory, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Tuscany, Italy.
  • Ognibene A; Chemical-Clinical Analysis Laboratory, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Tuscany, Italy.
Diagnostics (Basel) ; 14(5)2024 Mar 06.
Article in En | MEDLINE | ID: mdl-38473029
ABSTRACT
The Erythrocyte Sedimentation Rate (ESR) is a diagnostic estimator of systemic inflammation as a reflection of acute phase proteins circulating in the blood. The purpose of this manuscript is to evaluate the blood stability at room temperature (RT) and at 4 °C to avoid ESR diagnostic errors, as well as the accuracy of the VES-MATIC 5 analyzer. The ESR stability evaluation at RT for 24 h (4 h "T1", 6 h "T2", 8 h "T3", 10 h "T4", 24 h "T5") and at 4 °C (24 h, 36 h, 48 h) was carried out using 635 total samples, starting with T0 (2 h of venipuncture). For method comparison, 164 patients were analyzed using VES-MATIC 5 and then the Westergren reference method. The sample at RT is established by a significant gradual decrease in correlation R = 0.99 (T0 vs. T1), R = 0.97 (T0 vs. T2), R = 0.92 (T0 vs. T3), R = 0.87 (T0 vs. T4), and R = 0.40 (T0 vs. T5). The stability at 4 °C after 24 h, 36 h, and 48 h showed a regression of R = 0.99, R = 0.97, and R = 0.95, respectively. Therefore, ESR measurements on RT samples beyond 6 h after collection cannot be carried out, but the ESR can be measured until 36 h for samples stored at 4 °C. Moreover, the VES-MATIC 5 accuracy performance compared to the Westergren method (R = 0.96) is confirmed.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Diagnostics (Basel) Year: 2024 Document type: Article Affiliation country: Italia Country of publication: Suiza

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Diagnostics (Basel) Year: 2024 Document type: Article Affiliation country: Italia Country of publication: Suiza