Reliability of continuous vital sign monitoring in post-operative patients employing consumer-grade fitness trackers: A randomised pilot trial.
Helmer, Philipp; Hottenrott, Sebastian; Wienböker, Kathrin; Pryss, Rüdiger; Drosos, Vasileios; Seitz, Anna Katharina; Röder, Daniel; Jovanovic, Aleksandar; Brugger, Jürgen; Kranke, Peter; Meybohm, Patrick; Winkler, Bernd E; Sammeth, Michael.
Affiliation
  • Helmer P; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Hottenrott S; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Wienböker K; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Pryss R; Institute for Clinical Epidemiology and Biometry, University Hospital Würzburg, Würzburg, Germany.
  • Drosos V; Department of Thoracic and Cardiovascular Surgery, University Hospital Würzburg, Würzburg, Germany.
  • Seitz AK; Department of Urology and Pediatric Urology, University Hospital Würzburg, Würzburg, Germany.
  • Röder D; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Jovanovic A; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Brugger J; Department of Applied Sciences and Health, Coburg University of Applied Sciences and Art, Coburg, Germany.
  • Kranke P; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Meybohm P; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Winkler BE; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Sammeth M; Department of Anaesthesiology, Intensive Care, Emergency and Pain Medicine, University Hospital Würzburg, Würzburg, Germany.
Digit Health ; 10: 20552076241254026, 2024.
Article in En | MEDLINE | ID: mdl-38746874
ABSTRACT

Introduction:

Fitness trackers can provide continuous monitoring of vital signs and thus have the potential to become a complementary, mobile and effective tool for early detection of patient deterioration and post-operative complications.

Methods:

To evaluate potential implementations in acute care setting, we included 36 patients after moderate to major surgery in a recent randomised pilot trial to compare the performance of vital sign monitoring by three different fitness trackers (Apple Watch 7, Garmin Fenix 6pro and Withings ScanWatch) with established standard clinical monitors in post-anaesthesia care units and monitoring wards.

Results:

During a cumulative period of 56 days, a total of 53,197 heart rate (HR) measurements, as well as 12,219 measurements of the peripheral blood oxygen saturation (SpO2) and 28,954 respiratory rate (RR) measurements were collected by fitness trackers. Under real-world conditions, HR monitoring was accurate and reliable across all benchmarked devices (r = [0.95;0.98], p < 0.001; Bias = [-0.74 bpm;-0.01 bpm]; MAPE∼2%). However, the performance of SpO2 (r = [0.21;0.68]; p < 0.001; Bias = [-0.46%;-2.29%]; root-mean-square error = [2.82%;4.1%]) monitoring was substantially inferior. RR measurements could not be obtained for two of the devices, therefore exclusively the accuracy of the Garmin tracker could be evaluated (r = 0.28, p < 0.001; Bias = -1.46/min). Moreover, the time resolution of the vital sign measurements highly depends on the tracking device, ranging from 0.7 to 117.94 data points per hour.

Conclusion:

According to the results of the present study, tracker devices are generally reliable and accurate for HR monitoring, whereas SpO2 and RR measurements should be interpreted carefully, considering the clinical context of the respective patients.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Digit Health Year: 2024 Document type: Article Affiliation country: Alemania Country of publication: Estados Unidos

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Digit Health Year: 2024 Document type: Article Affiliation country: Alemania Country of publication: Estados Unidos