Your browser doesn't support javascript.
loading
Imaging of the vascular distribution of the outer ear using optical coherence tomography angiography for highly accurate positioning of a hearable sensor.
Hong, Juyeon; Seong, Daewoon; Kang, Dongwan; Kim, Hyunmo; Jang, Jeong Hun; Jeon, Mansik; Kim, Jeehyun.
Affiliation
  • Hong J; School of Electronic and Electrical Engineering, College of IT Engineering, Kyungpook National University, 80, Daehak-ro, Buk-gu, Daegu 41566, South Korea.
  • Seong D; School of Electronic and Electrical Engineering, College of IT Engineering, Kyungpook National University, 80, Daehak-ro, Buk-gu, Daegu 41566, South Korea.
  • Kang D; School of Electronic and Electrical Engineering, College of IT Engineering, Kyungpook National University, 80, Daehak-ro, Buk-gu, Daegu 41566, South Korea.
  • Kim H; School of Electronic and Electrical Engineering, College of IT Engineering, Kyungpook National University, 80, Daehak-ro, Buk-gu, Daegu 41566, South Korea.
  • Jang JH; Department of Otolaryngology, School of Medicine, Ajou University, 206, World cup-ro, Yeongtong-gu, Suwon 16499, South Korea.
  • Jeon M; School of Electronic and Electrical Engineering, College of IT Engineering, Kyungpook National University, 80, Daehak-ro, Buk-gu, Daegu 41566, South Korea.
  • Kim J; School of Electronic and Electrical Engineering, College of IT Engineering, Kyungpook National University, 80, Daehak-ro, Buk-gu, Daegu 41566, South Korea.
APL Bioeng ; 8(2): 026113, 2024 Jun.
Article in En | MEDLINE | ID: mdl-38799376
ABSTRACT
Novel hearable technology is securely and comfortably positioned within the ear canal minimizing inaccuracies caused by accessory movements during activities. Despite extensive research on hearable technologies within the outer ear, there is a lack of research in the field of vascular imaging and quantitative analysis in the outer ear in vivo, which is one of the crucial factors to select the appropriate sensor position. Therefore, in this paper, we introduced optical coherence tomography angiography (OCTA)-based qualitative and quantitative analyses to visualize the inner vasculature of the outer ear to acquire vascular maps for microvascular assessments in vivo. By generating maximum amplitude projection images from three-dimensional blood vascular volume, we identified variations of blood vessel signal caused by the different biological characteristics and curvature of the ear among individuals. The performance of micro-vascular mapping using the proposed method was validated through the comparison and analysis of individual vascular parameters using extracted 20 vascular-related variables. In addition, we extracted pulsatile blood flow signals, demonstrating its potential to provide photoplethysmographic signals and ear blood maps simultaneously. Therefore, our proposed OCTA-based method for ear vascular mapping successfully provides quantitative information about ear vasculature, which is potentially used for determining the position of system-on-chip sensors for health monitoring in hearable devices.

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: APL Bioeng Year: 2024 Document type: Article Affiliation country: Corea del Sur Country of publication: Estados Unidos

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: APL Bioeng Year: 2024 Document type: Article Affiliation country: Corea del Sur Country of publication: Estados Unidos