Correlation between vertebral bone mineral density and multi-level virtual non-calcium imaging parameters from dual-layer spectral detector computed tomography.
Yang, Yanhui; Hou, Jing; Niu, Yue; Zhang, Yi; Luo, Tao; Lu, Qiang; Fu, Yi; Wang, Yu; Yu, Xiaoping.
Affiliation
  • Yang Y; Department of Diagnostic Radiology, Graduate Collaborative Training Base of Hunan Cancer Hospital, Hengyang Medical School, University of South China, Hengyang, China.
  • Hou J; Department of Diagnostic Radiology, Hunan Cancer Hospital and the Affiliated Cancer Hospital of Xiangya School of Medicine, Central South University, Changsha, China.
  • Niu Y; Department of Diagnostic Radiology, Hunan Cancer Hospital and the Affiliated Cancer Hospital of Xiangya School of Medicine, Central South University, Changsha, China.
  • Zhang Y; Department of Diagnostic Radiology, Graduate Collaborative Training Base of Hunan Cancer Hospital, Hengyang Medical School, University of South China, Hengyang, China.
  • Luo T; Department of Diagnostic Radiology, Hunan Cancer Hospital and the Affiliated Cancer Hospital of Xiangya School of Medicine, Central South University, Changsha, China.
  • Lu Q; Department of Diagnostic Radiology, Graduate Collaborative Training Base of Hunan Cancer Hospital, Hengyang Medical School, University of South China, Hengyang, China.
  • Fu Y; Department of Diagnostic Radiology, Hunan Cancer Hospital and the Affiliated Cancer Hospital of Xiangya School of Medicine, Central South University, Changsha, China.
  • Wang Y; Department of Diagnostic Radiology, Hunan Cancer Hospital and the Affiliated Cancer Hospital of Xiangya School of Medicine, Central South University, Changsha, China.
  • Yu X; Department of Diagnostic Radiology, Hunan Cancer Hospital and the Affiliated Cancer Hospital of Xiangya School of Medicine, Central South University, Changsha, China.
Quant Imaging Med Surg ; 14(6): 3803-3815, 2024 Jun 01.
Article in En | MEDLINE | ID: mdl-38846313
ABSTRACT

Background:

Virtual non-calcium (VNCa) imaging based on dual-energy computed tomography (CT) plays an increasingly important role in diagnosing spinal diseases. However, the utility of VNCa technology in the measurement of vertebral bone mineral density (BMD) is limited, especially the VNCa CT value at multiple calcium suppression levels and the slope of VNCa curve. This retrospective cross-sectional study aimed to explore the correlation between vertebral BMD and new VNCa parameters from dual-layer spectral detector CT.

Methods:

The dual-layer spectral detector CT and quantitative CT (QCT) data of 4 hydroxyapatite (HAP) inserts and 667 vertebrae of 234 patients (132 male and 102 female) who visited a university teaching hospital between April and May 2023 were retrospectively analyzed. The BMD values of 3 vertebrae (T12, L1, and L2) and inserts were measured using QCT, defined as QCT-BMD. The VNCa CT values and the slope λ of the VNCa attenuation curve of vertebrae and inserts were recorded. The correlations between VNCa parameters (VNCa CT value, slope λ) and QCT-BMD were analyzed.

Results:

For the vertebrae, the correlation coefficient ranged from -0.904 to 0.712 (all P<0.05). As the calcium suppression index (CaSI) increased, the correlation degree exhibited a decrease first and then increased, with the best correlation (r=-0.904, P<0.001) observed at the index of 25%. In contrast, the correlation coefficient for the inserts remained relatively stable (r=-0.899 to -1, all P<0.05). For the vertebrae, the values of 3 slopes λ (λ1, λ2, and λ3) derived from the VNCa attenuation curve were 6.50±1.99, 3.75±1.15, and 2.04±0.62, respectively. Regarding the inserts, the λ1, λ2, and λ3 values were 11.56 [interquartile range (IQR) 2.40-22.62], 6.68 (IQR 1.39-13.49), and 3.63 (IQR 0.75-7.8), respectively. For the vertebrae, all 3 correlation coefficients between 3 slopes λ and QCT-BMD were 0.956 (all P<0.05). For the inserts, the 3 correlation coefficients were 0.996, 0.998, and 1 (all P<0.05), respectively.

Conclusions:

A promising correlation was detected between VNCa CT parameters and QCT-BMD in vertebrae, warranting further investigation to explore the possibility of VNCa imaging to assess BMD.